Solar8000 ART and CVP numerics¶
Arterial and central venous pressure values from the Solar8000 monitor.
Stub
Imported from the time-source table; the full page is still to be written.
| Name in the dataset | Solar8000/ART_MBP |
| Type | numeric |
| Device | Solar8000 monitor |
| Stamped by | Recorder PC clock |
| Unit | mmHg |
Measurement space¶
Time side (after the per-case delay label): the true time lies between -8.4 s and -0.055 s from the stamp: the delay range plus the interval one value describes.
Time¶
From UTC to the stored time stamp¶
Derived from this signal's timing table; a written chain is still to come.
The clock path first (UTC to the clock that stamps or spaces the samples), then the signal's own path to its stored time. Each bar: the time shift or error the step introduces (pale) and what is left after the atlas's correction (solid), on a log scale; dashed when estimated or assumed.
The reference every chain starts from.
Windows time synchronisation to the hospital's time server every 8–24 h (assumed), drifting a few ppm in between. Not observable in the release: dates are shifted to 2100. No step of 0.5 s or more was found.
Delay or clock offset to the stamp; removed by the per-case delay label.
Stored time stamp
The width of the interval one value or sample stands for.
Timing¶
Against the working reference. Delay: from the end of the interval a value describes to its stamp. Interval: how wide that interval is.
| Form | Delay | Interval | Clock domain | Status | Evidence |
|---|---|---|---|---|---|
| after the per-case delay label (default) | 55 ms – 400 ms | 3 s – 8 s | Solar8000 numerics | measured | numeric_delay label SE p10-p90: ART 0.06-0.35 s, CVP 0.11-0.40 s (fixed within a case) |
| as stamped | 300 ms – 4.7 s | 3 s – 8 s | Solar8000 numerics | measured | delay from window end: ART 0.3-2.5 s, CVP 3.1-4.7 s; windows 3-6 s + 2 s poll |
Combine it with other signals in the synchronisation planner.
Delays¶
- Mean arterial pressure (1.7 s (p10–p90 1.0 to 2.5)): EWMA ¼ per beat; systolic (median over 3 s) and diastolic (EWMA ½) are about 0.5 s shorter.
- CVP (3.8 s (p10–p90 3.1 to 4.7)): Waveform mean over 4 s; the slowest Solar8000 numeric.
Value¶
Its value side is still to be described.
Parameters¶
| Parameter | Value | Status | From |
|---|---|---|---|
| Solar8000 ART mean delay | 1.7 s (p10–p90 1.0 to 2.5) | measured | Numeric delays are fixed within a case, different between cases |
| Solar8000 CVP delay | 3.8 s (p10–p90 3.1 to 4.7) | measured | Numeric delays are fixed within a case, different between cases |
Observations¶
| Observation | Status | Summary |
|---|---|---|
| Monitor numerics describe the past, never the future | established | Every monitor numeric describes a window of 3–30 s or more that ended 1–9 s before its stamp. None is stamped before its data exist, so the stamps already respect the availability rule. |
| Numeric delays are fixed within a case but differ between cases | established | A numeric's delay does not drift through a recording, but 40–70% of its variance lies between cases. One delay per case describes it; one delay for the whole dataset is wrong by up to about 2 s. |
Studies¶
| Study | Date | Status | Summary |
|---|---|---|---|
| The SNUADC clock through the Solar8000 numerics, whole release | 2026-10-19 | current | On the SNUADC's own sample count, the delay from the waveform to the PC-stamped Solar8000 heart rate and mean arterial pressure ramps at the SNUADC's drift against the PC, measured without its stamps or resets. In recordings with whole reset intervals it agrees with them (−63 ppm; median difference +1.5 ppm, 581 recordings). It is too noisy per recording, and in recordings under about 4 h its rate is biased toward zero, so it cannot yet speak for recordings without resets. |
| Value resolution of every signal | 2026-10-19 | draft | How finely every waveform and numeric track in the release records its values. The SNUADC waveforms are 10-bit codes, though VitalRecorder's documentation describes the SNUADC as 12-bit; Primus waveforms use about 10 bits of 12; BIS EEG is a 16-bit grid of 0.05 µV. Numerics are 32-bit floats on decimal grids of 1, 0.1, 0.01 or 0.001, one per track; settings sit on coarser grids of 10 or 100. |
| Numeric delays are fixed within a case, different between cases | 2026-10-12 | current | With each numeric's algorithm held fixed, delays were fitted for every case and every 30-minute block. They do not drift within a case, but 40–70% of their variance lies between cases, so a single population delay is wrong by up to 2 s for HR and 1 s for the others. Each case gets its own delay label. |
| How the monitors compute their numbers | 2026-10-12 | current | Recomputing each monitor numeric from its own waveform finds a short, simple algorithm for every one (a median, an exponential average or a time mean over 3–30 s) and a delay of 2–9 s behind the newest data it uses. No value is stamped before its data exist. |
Worked examples¶
| Example | Side | Status | Summary |
|---|---|---|---|
| Cuff pressure against the arterial line | both | draft | 12,238 cuff readings against the arterial line at the cuff's own moment, in 2,283 patients: no bias over the cohort at any pressure, but each patient's cuff reads a constant 10 mmHg (SD) above or below their line, and half the arterial readings below 65 mmHg are not below 65 by cuff. |
| Minutes of hypotension per patient, from the monitor and from the waveform | both | draft | Minutes with MAP below 65 mmHg in 3,476 recordings, from the monitor's numeric and from the arterial waveform's beats: the two routes to one pressure differ by several mmHg per patient, which moves half of all patients' minutes by 5 or more and a fifth of their labels, and the numeric's averaging hides short dips in every patient. |
| Post-induction hypotension, from the charted start and from the devices | time | draft | The charted anaesthesia start comes a median 14 minutes before the patient is induced; measured from it, "hypotension within 20 minutes of induction" is found in 27% of 2,913 patients, against 51% from the devices' induction, and the label changes for a quarter of patients. |
| Pulse arrival time against blood pressure, before and after synchronisation | time | draft | In 3,270 recordings, PAT tracks systolic pressure only once each pulse is paired with its own beat: the hourly correlation moves from a mean of −0.04 to −0.21 (88% of recordings negative), reproducing Ruffolo et al. 2025 in a second dataset; paired naively, PAT is a heart-rate signal. |
Tags: stub