SNUADC channels are read one after another

The SNUADC firmware reads its 8 channels in turn, about 0.1 ms each, so within one sample the last channel is read up to about 0.8 ms after the first. The channel order is not known.

Status likely: consistent, with a known gap.
Devices SNUADC
Variables ECG lead II (SNUADC), Arterial pressure (SNUADC), Plethysmogram (SNUADC)

What it means

A fixed sub-millisecond skew between SNUADC channels: a floor on same-clock alignment such as ECG to arterial pressure. A monitor line seen in both ART and PLETH puts ART about 0.3–0.4 ms ahead.

Evidence

Study Date Status Summary
Monitor clocks in the waveforms' spectral lines 2026-10-17 open Differences of the raw converter counts show lines from the monitor's analogue outputs: 240 Hz in arterial pressure and plethysmogram, and a 30 Hz family in the plethysmogram alone. Each measures a monitor clock against SNUADC.
The SNUADC firmware 2026-10-13 current The public SNUADC firmware explains the board's timing: it sends no time stamps, samples on its 16 MHz crystal, loses time rather than bunching samples after a stall, reads its channels one after another, and truncates its values. A predicted fault at the microsecond counter's wrap was not found in the data.