The EVENT track repeats the clinical table¶
In 411 random cases the recorder EVENT track's four entries equal the clinical table's case start, operation start, operation end and case end exactly: the release generated them from the table.
| Status | established: shown on many cases, with a control. |
| Devices | Vital Recorder, Clinical records |
| Variables | Anaesthesia record events, Case end |
What it means¶
The EVENT track is not an independent record of when those events happened.
Evidence¶
| Study | Date | Status | Summary |
|---|---|---|---|
| Closing loops in the timing graph | 2026-10-13 | current | Where two independent routes reach the same clock, the offsets around the loop must sum to zero. Three measured loops close within their noise; the BIS loop closes up to a constant latency; the proposed EMR loop is not a loop at all. |