Time between the Primus's small corrections.
| Value | about 7,230 s (2.0 h; p90 twice that) |
| Status | measured: Measured in a study. |
| From | When each device's clock is corrected, in every recording |
| About | Primus against the recorder PC |
| Date | 2026-10-19 |
Notes¶
The whole release: 7,230 s or twice that, 8,077 corrections in 4,529 recordings; the AWP pauses recur at the same interval (The Primus AWP stops for about 21 s every 7,230 s). The pilot (How often each device is corrected) gave the same.
Used in¶
- correction period of Primus against the recorder PC