Solar8000 numerics against SNUADC

The Solar8000's numbers carry no clock of their own in the release; their delay behind the SNUADC waveforms they summarise is fixed within a case, which bounds any drift of the monitor's timing.

Clock Solar8000 clock
Measured against SNUADC sample clock
Modes all
Status measured
Drift 0 s/h median (p10–p90 about ±0.5 s/h)

Correction

None needed within a case; one delay per case (the numeric_delay label) puts each numeric on the SNUADC timeline.

Parameters

Parameter Value Status From
Drift of Solar8000 numeric delays within a case 0 s/h median (p10–p90 about ±0.5 s/h) measured Numeric delays are fixed within a case, different between cases

Evidence

Study Date Status Summary
Numeric delays are fixed within a case, different between cases 2026-10-12 current With each numeric's algorithm held fixed, delays were fitted for every case and every 30-minute block. They do not drift within a case, but 40–70% of their variance lies between cases, so a single population delay is wrong by up to 2 s for HR and 1 s for the others. Each case gets its own delay label.