Vital Recorder

The recording software that collects every device's data and writes the .vital files.

Type software
Clocks Recorder PC clock

Description

Vital Recorder runs on a PC in each operating room. It stamps numeric records on arrival and stamps waveform records from its model of each device's timing, re-synchronising that model when it drifts too far.

Device documentation

Observations

Observation Status Summary
Device clock drift depends on the setup established The drift of the SNUADC's and the BIS's timestamps against the recorder PC is largely set by the setup: knowing a recording's PC key (the serial ports of its four core devices) puts the SNUADC's drift within 3 ppm of its setup's median in 97% of cases and the BIS's in 83%, against 34% and 17% for the release's median. On the whole release's correction intervals the same test gives 71% and 60% (Identifying the recording setup of every case).
Eight recordings store whole messages twice established In 8 recordings 15–48% of the waveform records are byte-for-byte copies of the record before them, with the same timestamp: case 15, case 364, case 1946, case 3865, case 4857, case 5414, case 5595 and case 5619. SNUADC, Primus and BIS are affected equally, so the recorder, not a device, wrote them twice.
Every case starts at 2100-01-01 00:00 UTC established The release shifts each case to the same start, removing the date and the time of day. Within a case all times keep their spacing.
Five records out of order at every SNUADC reset established At each step-mode reset the five records after it are written in the wrong order: two continue the old timing and three start the new, interleaved. Read in file order, or sorted by stamp, one second of signal is scrambled.
Recording pauses of about 20 s established In 29% of files every device jumps forward together, by a median 19.6 s: the recorder paused. Data are missing there, and the stamps show it honestly. The pauses run on a 5-hour timer: consecutive pauses in a recording are 5 h apart (plus the pause) in 86% of cases, and the first comes at any time after the recording starts.
SNUADC sends no time; the recorder stamps by count established The SNUADC board streams samples with no time stamps. Every SNUADC time in VitalDB is the recorder's model of the board's clock: the sample count at the nominal rate, pulled back when it drifts too far.
SNUADC stamps run −0.95 ppm against the sample count established Between corrections each 0.2 s record stamp increment is about 0.19 µs short: stamps run −0.95 ppm from the sample count in 88% of files (both SNUADC modes). It is arithmetic, not clock drift. The other 12% sit between −0.90 and −0.55 ppm.
Solar8000 heart rate comes one poll early or late established Heart rate reaches the recorder about 1.9 s or about 4.2 s after its window ends, one 2 s poll apart. The group belongs to the recording PC and stays fixed through a case.
The EVENT track repeats the clinical table established In 411 random cases the recorder EVENT track's four entries equal the clinical table's case start, operation start, operation end and case end exactly: the release generated them from the table.
Time zero depends on the tracks loaded established A file's time zero is its first record of any track. Loading only some tracks moves the zero to their first record: seconds later, and up to hours.
In 34 files the SNUADC records overlap throughout open In 34 files the SNUADC records step back over their neighbours far more often than resets explain: in 20 of them 4–95% of all record junctions overlap, by up to about 100 ms. The reset-order rule matches there by coincidence.
In some recordings two unrelated streams are interleaved in one track open In 7 recordings 64–95% of the records alternate between two streams: each record continues the one two before it, not the one before, and the two streams do not match each other. SNUADC and BIS in case 2219, case 2379, case 2640, case 4836, case 5327 and case 5617; SNUADC in case 1939. In case 94, case 945, case 1132, case 3195 and case 5700 4–7% of SNUADC records interleave.

Studies

Study Date Status Summary
Gaps and unbroken segments in every waveform 2026-10-19 draft Every gap and every unbroken segment in the 6,388 recordings, per waveform. The curves are not smooth: a cliff at about 18 s in every device marks the recorder's pauses, SNUADC segments stop at 5 hours, Primus segments at about 2.0 hours and BIS segments at about 2.3 hours (each device's correction cycle), and BIS's short gaps grow with the time since the last one (clock resynchronisations, not lost data).
Message sizes and the steps between records 2026-10-19 draft Every record in every waveform: SNUADC writes messages of 100 samples (0.2 s), Primus 30 (0.48 s), BIS 16 (0.125 s), CardioQ no fixed size. Backward steps outnumber gaps by about a hundred to one and come from a few recordings with duplicated or interleaved records. Primus gaps are whole lost messages; most SNUADC and BIS gaps are not.
The SNUADC clock through the Solar8000 numerics, whole release 2026-10-19 current On the SNUADC's own sample count, the delay from the waveform to the PC-stamped Solar8000 heart rate and mean arterial pressure ramps at the SNUADC's drift against the PC, measured without its stamps or resets. In recordings with whole reset intervals it agrees with them (−63 ppm; median difference +1.5 ppm, 581 recordings). It is too noisy per recording, and in recordings under about 4 h its rate is biased toward zero, so it cannot yet speak for recordings without resets.
What each break in the stamps is 2026-10-19 draft Every step between waveform records in the release, sorted by mechanism: written twice, a SNUADC clock reset, interleaved streams, a pause shared by every device, whole messages, less than a message, or a fraction of a message beyond whole ones. A continuity test with controls shows which classes lose data: shared pauses and most longer SNUADC gaps do; steps under one message, and SNUADC's lone one-message steps, mostly do not.
Identifying the recording setup of every case 2026-10-18 open Recognising which equipment made each recording, from its serial ports and, where those fall short, from clock fingerprints, so that per-setup values can be used for every case.
Paced ECG as an independent clock 2026-10-17 current Fixed-rate pacing gives a clock the recording chain cannot touch. Timing paced beats on three timelines shows that each SNUADC step-mode correction loses about 18 ms of signal, and that nothing else breaks the sample stream.
Recorder PC clock steps and recording pauses 2026-10-17 current A step of the recorder PC clock would move every device's stamps at once. Searching all 6,388 files for such simultaneous jumps found none of 0.5 s or more; what moves everything together is recording pauses.
Records out of order at SNUADC resets, and where the lost time goes 2026-10-17 current At every SNUADC reset the five records after it are written out of order: two continue the old timing and three start the new. Put back in order, the waveform runs on with no samples missing; the 18–20 ms lost at each reset goes missing inside the last old-timing record, and an extra zero sample follows.
SNUADC stamp arithmetic 2026-10-17 current Between corrections the SNUADC block stamps run −0.95 ppm against the sample count: each 0.2 s increment is about 0.19 µs short. This, not clock drift, is the "about −1 ppm" seen in frequent-correction mode.
The vitaldb package's loading conventions 2026-10-17 current The vitaldb Python package sets time zero from the tracks loaded, places waveform blocks by rounding their stamps up to a sample, and lets overlapping blocks overwrite each other. Each choice moves or loses samples.
Time zeros and loaded subsets 2026-10-16 current A VitalDB time "from the start" depends on which tracks were loaded: loading a subset moves the zero to that subset's first record, seconds or hours later. The arrhythmia database's annotations count from the full file's zero.
Two heart-rate delay groups, one poll apart 2026-10-13 current Solar8000 heart rate reaches the recorder either about 1.9 s or about 4.2 s after its window ends: one 2 s poll apart. The group belongs to the recording PC, stays fixed through a case, and on some PCs varies from case to case.
Closing loops in the timing graph 2026-10-13 current Where two independent routes reach the same clock, the offsets around the loop must sum to zero. Three measured loops close within their noise; the BIS loop closes up to a constant latency; the proposed EMR loop is not a loop at all.
The SNUADC firmware 2026-10-13 current The public SNUADC firmware explains the board's timing: it sends no time stamps, samples on its 16 MHz crystal, loses time rather than bunching samples after a stall, reads its channels one after another, and truncates its values. A predicted fault at the microsecond counter's wrap was not found in the data.
Recovering the time of day 2026-10-12 current The release shifts every case to start at 2100-01-01 00:00 UTC, removing the date and the time of day. The time of day (Korean local time) can be recovered to the minute from the admission date and the seconds of the EMR times.
Anaesthesia record times are rounded up 2026-10-12 current Anaesthesia start and end entered on a 5-minute mark sit 90–290 s later, against device markers, than entries typed to the minute. Rounding up to the next 5 minutes fits every one of 11 markers better than rounding to the nearest or down.
The stamp sawtooth model for SNUADC and BIS 2026-10-12 current Fitting each track's record stamps against its sample count as a drift with corrections (a sawtooth) puts SNUADC and BIS on one timeline without any artefact. Checked against shared bursts, the EEG–ECG timing error falls from about 790 ms on the recorded stamps to 33 ms with one constant per SNUADC mode in the pilot, and 42 ms over the whole release.
Clock corrections in every file 2026-10-11 current A scan of all 6,388 files for the jumps where the recorder re-synchronises a device's stamps. Each device type has its own correction signature; only SNUADC and BIS correct clearly enough to measure their drift.
Physical units, shared drift and time order 2026-10-11 current Device drift rates cluster into units: about 8 SNUADC boards and 8–18 BIS monitors. Part of the drift is shared by both devices of a case and changes between cases: the recorder PC's own rate. Case order carries no time information.

References

  • Lee H-C, Park Y, Yoon SB, Yang SM, Park D, Jung C-W. VitalDB, a high-fidelity multi-parameter vital signs database in surgical patients. Sci Data. 2022;9:279. ↗
  • Lee H-C, Jung C-W. Vital Recorder – a free research tool for automatic recording of high-resolution time-synchronised physiological data from multiple anaesthesia devices. Sci Rep. 2018;8:1527. ↗