Skip to content

SNUADC frequent-correction steps

Forward steps of the SNUADC stamps in frequent-correction mode.

Value +5 to +15 ms (median +10), about one an hour (median 1.1 per hour per recording)
Status measured: Measured in a study.
From Clock corrections in every file
About SNUADC (frequent-correction mode) against the recorder PC

Notes

The rate is from the correction scan's step counts (Clock corrections in every file): 635 small corrections in 448 hours of the 120 recordings in frequent-correction mode, a median of 1.1 per hour per recording (p10–p90 0.5–4.1), that is one every 15 minutes to 2 hours. An earlier figure of 0.2 per hour was not traceable to any result and has been replaced.