Five records out of order at every SNUADC reset

At each step-mode reset the five records after it are written in the wrong order: two continue the old timing and three start the new, interleaved. Read in file order, or sorted by stamp, one second of signal is scrambled.

Status established: shown on many cases, with a control.
Clocks SNUADC sample clock
Devices SNUADC, Vital Recorder
Variables ECG lead II (SNUADC), Arterial pressure (SNUADC), Plethysmogram (SNUADC)

What it means

Any reader that keeps file order or sorts by stamp (the vitaldb package included) scrambles about 1 s of every SNUADC channel at each reset, roughly every two hours of step-mode recording.

What is done about it

The TimeNet import puts the records back in sampling order by default (VITALDB_RESET_ORDER), and the clock fits use the same order; a fit and a build in different orders are not combined.

Evidence

Study Date Status Summary
Records out of order at SNUADC resets, and where the lost time goes 2026-10-17 current At every SNUADC reset the five records after it are written out of order: two continue the old timing and three start the new. Put back in order, the waveform runs on with no samples missing; the 18–20 ms lost at each reset goes missing inside the last old-timing record, and an extra zero sample follows.
Clock corrections in every file 2026-10-11 current A scan of all 6,388 files for the jumps where the recorder re-synchronises a device's stamps. Each device type has its own correction signature; only SNUADC and BIS correct clearly enough to measure their drift.