Monitor clocks in the waveforms' spectral lines¶
Differences of the raw converter counts show lines from the monitor's analogue outputs: 240 Hz in arterial pressure and plethysmogram, and a 30 Hz family in the plethysmogram alone. Each measures a monitor clock against SNUADC.
| Date | 2026-10-17 |
| Status | open |
| Methods | Clocks carried in the signal |
| Clocks | Monitor analogue output clock, Pulse oximeter frame clock, SNUADC sample clock |
| Devices | Solar8000 monitor, SNUADC |
| Variables | Arterial pressure (SNUADC), Plethysmogram (SNUADC), ECG lead II (SNUADC) |
Data¶
6 short-to-medium files (1.4–2.6 h), chosen at random; work of the R-peak timing session, consolidated here.
Method¶
Integer counts were recovered from each track and the spectrum of their sample-to-sample changes searched for lines. The 240 Hz line was demodulated in 10 s windows on the global sample index; its frequency came from a zero-padded FFT with parabolic peak interpolation. The phase difference of ART and PLETH at that line gave their skew.
Results¶
- 240 Hz in PLETH and ART (weakly in ECG in 3 cases): −25.4 to −28.7 ppm against the SNUADC.
- A 30 Hz family (30, 60, 90, 150 Hz) in PLETH only: −30 to −73 ppm in frequency, that is +30 to +73 ppm in period, matching the oximeter frame clock from fiducial locking (The PPG's 30 Hz frame clock).
- ART leads PLETH by 0.31–0.39 ms at 240 Hz in every case, consistent with the converter reading its channels one after another.
- Line frequencies group files by device: two of the six share every line and are probably the same monitor and converter.
- No frame-slip sawtooth was seen in the ART→PLETH lag of these six files, which conflicts with The PPG sawtooth and its correction; a few slips per short file barely show in a jump histogram, and many stretches have a paused sawtooth. To settle by running these files through physiosync.
| Parameter | Value | Status | From |
|---|---|---|---|
| Monitor 240 Hz line against SNUADC | −25.4 to −28.7 ppm (6 cases) | measured | Monitor clocks in the waveforms' spectral lines |
Discussion¶
A third, independent clock (the monitor's output stage) measured from the waveforms alone, and a blind way to group recordings by device.
Observations it supports¶
| Observation | Status | Summary |
|---|---|---|
| The plethysmogram is built in 30 Hz frames | established | The SNUADC PLETH waveform comes from the oximeter in 33.3 ms frames, smoothed to 500 Hz. Its content changes once per frame, on the oximeter's own clock, about 0.4 s after the finger. |
| SNUADC channels are read one after another | likely | The SNUADC firmware reads its 8 channels in turn, about 0.1 ms each, so within one sample the last channel is read up to about 0.8 ms after the first. The channel order is not known. |
Limitations¶
Six files.
Open questions¶
- Monitor clock lines in six files only (next step). The monitor clock lines were seen in six files; running these files through physiosync, and then a larger set, would settle whether they are general.