SNUADC (step mode) against the recorder PC
In step mode the SNUADC stamps drift by tens of ppm and are pulled back by about half a second whenever the drift reaches that size, about every two hours.
Correction
Triggered by drift, not by a timer: once the stamps are about 0.5 s ahead, the recorder pulls the track back.
The stamps show only −0.480 s of the true correction of about 0.498 s, because about 18 ms of signal is never
sampled (Signal lost at a SNUADC step-mode correction). In all recordings (When each device's clock is corrected, in every recording): 3,417
resets in 2,540 recordings, a median 2.08 h apart; the drift count usually starts with the recording (the first
reset falls at 0.99 × the time 0.49 s of drift takes, interquartile 0.87–1.01), though not reliably enough to
time drift from a single reset. Rates are relative to the recorder PC.
Parameters
Evidence
| Study |
Date |
Status |
Summary |
| Paced ECG as an independent clock |
2026-10-17 |
current |
Fixed-rate pacing gives a clock the recording chain cannot touch. Timing paced beats on three timelines shows that each SNUADC step-mode correction loses about 18 ms of signal, and that nothing else breaks the sample stream. |
| The stamp sawtooth model for SNUADC and BIS |
2026-10-12 |
current |
Fitting each track's record stamps against its sample count as a drift with corrections (a sawtooth) puts SNUADC and BIS on one timeline without any artefact. Checked against shared bursts, the EEG–ECG timing error falls from about 790 ms on the recorded stamps to 33 ms with one constant per SNUADC mode in the pilot, and 42 ms over the whole release. |