Parameters

One number the atlas relies on, with its uncertainty and where it came from.

Parameter Value Status From
Adopted mean pulse arrival time 270 ms (plausible range 250–290 ms) assumed The PPG hardware delay and the adopted pulse arrival time
Arterial line access rate 1.1 per hour (8,245 in 7,507 h); occlusions 0.4 per hour measured Arterial line accesses in the pressure waveform
BIS correction interval 8,192 s (2 h 16.5 min), ± 20 s in 95% of intervals measured When each device's clock is corrected, in every recording
BIS correction size 0.17 s median measured When each device's clock is corrected, in every recording
BIS drift against SNUADC −72 ppm (IQR −82 to −54) measured Shared bursts between the ECG and the EEG
BIS drift against the recorder PC 25 ppm (p10–p90 12 to 50; 5th–95th 7.3–51.1 ppm; census fit of 4,728 recordings median 21.1 ppm) measured The stamp sawtooth model for SNUADC and BIS
BIS index averaging window 16 s in 48% of good fits (the 15 s setting); 32–61 s in 40%; half the weight older than 30 s measured How the BIS index trails its EEG
BIS index delay 3 s (p10–p90 3 to 4) measured How the BIS index trails its EEG
Blood gas timestamp after the draw 7.2 min (p10–p90 3.2 to 12.2; two modes, 2–6 and 8–12 min) measured What the lab timestamp marks
CardioQ re-anchoring interval at most about 300 s; about 40 re-anchors per case measured —
CardioQ stamps behind SNUADC 0.4–4.4 s (p5–p95); matched-beat MAD 0.24 s (24 cases) measured Rare devices against SNUADC
Central lab timestamp after the draw about 8–24 min (median about 14; weak, mostly chance matches) estimated What the lab timestamp marks
Drift of Solar8000 numeric delays within a case 0 s/h median (p10–p90 about ±0.5 s/h) measured Numeric delays are fixed within a case, different between cases
EEG–ECG constant after the stamp model −804 ms in step mode (1,526 recordings); −660 ms in frequent-correction mode (78) measured The stamp sawtooth model for SNUADC and BIS
EEG–ECG timing error after the stamp model 42 ms median, p90 150 ms (17 ms and 79 ms with the recording's own bursts) measured The stamp sawtooth model for SNUADC and BIS
EMR 5-minute entries, rounding (t − 300 s, t]; estimate t − 150 s (SD 87 s); observed shift against minute entries −89 to −293 s measured Anaesthesia record times are rounded up
Largest recorder PC clock step no step of 0.5 s or more in 6,388 files (19,006 h) measured Recorder PC clock steps and recording pauses
Monitor 240 Hz line against SNUADC −25.4 to −28.7 ppm (6 cases) measured Monitor clocks in the waveforms' spectral lines
Oximeter frame clock against SNUADC +7 to +92 ppm by setup (medians over 42 PC keys); within 3 ppm of its setup's median in 89% of recordings measured The PPG's 30 Hz frame clock
PPG hardware delay 385 ms, one global value (stored R-to-foot delay 655 ms minus an adopted 270 ms); setups differ by under ±15 ms measured One device delay for the PLETH across the release
PPG sawtooth rate 18–52 ppm by setup measured The PPG sawtooth and its correction
PPG slip size 32.3 ms (IQR 30.4 to 34.8) measured The PPG sawtooth and its correction
Pacing interval agreement between stretches 0.05–0.1 ms measured Paced ECG as an independent clock
Primus ETCO₂ delay 1.55 s (p10–p90 0.75 to 3.4) measured Numeric delays are fixed within a case, different between cases
Primus PIP delay 4.6 s (p10–p90 3.65 to 6.5) measured Numeric delays are fixed within a case, different between cases
Primus RR delay 4.3 s (p10–p90 1.6 to 7.6) measured Numeric delays are fixed within a case, different between cases
Primus correction interval about 7,230 s (2.0 h; p90 twice that) measured When each device's clock is corrected, in every recording
Primus correction size whole 16 ms samples either way; median −16 ms (one sample back), p10–p90 −144 to +144 ms measured When each device's clock is corrected, in every recording
Primus drift against the recorder PC, from breath matching 104 ppm (p10–p90 28 to 184; 178 recordings; whole-recording lines give +87 ppm median in 1,060 (p10–p90 −49 to +229)) measured The Primus ventilator against SNUADC, by breaths
Primus drift implied by its corrections about 1–11 ppm derived How often each device is corrected
Primus drift used in the delay fits 115 ppm assumed —
Recorder PC drift against UTC 0.1–10 ppm (assumed) assumed —
Recorder PC offset from UTC 3 ms to 0.9 s (assumed) assumed —
Recorder PC synchronisation interval 8–24 h (assumed) assumed —
Recording pause 19.6 s (p10–p90 15.7 to 29.8; 2,025 pauses in 1,846 files (29%)) measured Recorder PC clock steps and recording pauses
SNUADC frequent-correction steps +5 to +15 ms (median +10), about one an hour (median 1.1 per hour per recording) measured Clock corrections in every file
SNUADC stamp arithmetic −0.954 ppm (n = 6321; census of 6,321 files (17 October): median −0.9535 ppm; 89.5% of recording hours within 0.05 ppm of −0.954; 747 files (12%) between −0.90 and −0.55 ppm) measured SNUADC stamp arithmetic
SNUADC start after the file's first record 1.56 s (5–95%: 0.44–4.5 s; 99th percentile 12 s; maximum 10,879 s) measured The vitaldb package's loading conventions
SNUADC step-mode correction interval 2.08 h (5th–95th 1.65–3.57 h between resets; per-recording means p10–p90 1.68–3.07 h) measured The stamp sawtooth model for SNUADC and BIS
SNUADC step-mode correction size about 0.498 s true (paced: 0.4981–0.4988 s); −0.480 s on the stamps measured Paced ECG as an independent clock
SNUADC step-mode drift −63 ppm (p10–p90 −79 to −44; 5th–95th −80.3 to −36.9 ppm; medians of −46 to −74 ppm by serial port) measured The stamp sawtooth model for SNUADC and BIS
Signal lost at a SNUADC step-mode correction 17.7 ms (range 17.2 to 18.2; n = 3; 3 corrections in 2 paced cases; ECG–EEG bursts give 13.0 ± 1.7 ms) measured Paced ECG as an independent clock
Solar8000 ART mean delay 1.7 s (p10–p90 1.0 to 2.5) measured Numeric delays are fixed within a case, different between cases
Solar8000 CVP delay 3.8 s (p10–p90 3.1 to 4.7) measured Numeric delays are fixed within a case, different between cases
Solar8000 HR delay, early group 1.92 s (± 0.51 SD; 36% of classified cases) measured Two heart-rate delay groups, one poll apart
Solar8000 HR delay, late group 4.2 s (± 0.65 SD; 64% of classified cases) measured Two heart-rate delay groups, one poll apart
Solar8000 HR window about 6 s (best 6–8 s in both delay groups) measured How the monitors compute their numbers
SpO₂ onset latency about 4 s to the start of the dip, 8 s to its deepest; window 10–20 s (estimated) measured SpO₂ and pulse rate latency from step events
Stamp shift across a recording gap 127–149 ms early (4 gaps, 2 cases; modulo one pacing period) measured Paced ECG as an independent clock