BIS against SNUADC
Measured directly from bursts seen in both the ECG and the EEG, independent of the recorder: −72 ppm (median of 191 benchmark cases), and a constant offset of about −0.8 s left after both stamp models. Per case the slope matches the difference of the two drifts against the PC (Shared bursts between the ECG and the EEG); the release medians of those drifts (+25 and −63 ppm) differ by 88 ppm, being medians over different sets of recordings.
Correction
Both devices' own corrections: SNUADC on its drift threshold, BIS on its 8,192 s timer.
Parameters
Evidence
| Study |
Date |
Status |
Summary |
| Shared bursts between the ECG and the EEG |
2026-10-12 |
current |
Electrocautery and muscle bursts appear in both the SNUADC ECG and the BIS EEG. Cross-correlating their envelopes gives the offset between the two devices' stamps every 10 minutes: about −0.8 s, in a sawtooth that follows both clocks' corrections. |
| The stamp sawtooth model for SNUADC and BIS |
2026-10-12 |
current |
Fitting each track's record stamps against its sample count as a drift with corrections (a sawtooth) puts SNUADC and BIS on one timeline without any artefact. Checked against shared bursts, the EEG–ECG timing error falls from about 790 ms on the recorded stamps to 33 ms with one constant per SNUADC mode in the pilot, and 42 ms over the whole release. |