Recording pause¶
Length of the pauses in which every device jumps forward together.
| Value | 19.6 s (p10–p90 15.7 to 29.8; 2,025 pauses in 1,846 files (29%)) |
| Status | measured: Measured in a study. |
| From | Recorder PC clock steps and recording pauses |
| About | Recorder PC clock |
Notes¶
The gap survey of every waveform (Gaps and unbroken segments in every waveform, 19 October) puts the same pauses as a cliff in the gap distribution of every SNUADC, Primus and BIS track: ECG II has 1,968 gaps of 15–25 s, median 17.9 s. The two medians use different definitions: the survey counts gaps in one track (from the end of one record to the start of the next), while this parameter counts moments when every device jumps together.