Recording pause

Length of the pauses in which every device jumps forward together.

Value 19.6 s (p10–p90 15.7 to 29.8; 2,025 pauses in 1,846 files (29%))
Status measured: Measured in a study.
From Recorder PC clock steps and recording pauses
About Recorder PC clock

Notes

The gap survey of every waveform (Gaps and unbroken segments in every waveform, 19 October) puts the same pauses as a cliff in the gap distribution of every SNUADC, Primus and BIS track: ECG II has 1,968 gaps of 15–25 s, median 17.9 s. The two medians use different definitions: the survey counts gaps in one track (from the end of one record to the start of the next), while this parameter counts moments when every device jumps together.