Skip to content

EEG–ECG timing error after the stamp model

EEG–ECG timing error per 10-minute window with the stamp model and the SNUADC mode's constant.

Value 42 ms median, p90 150 ms (17 ms and 79 ms with the recording's own bursts)
Status measured: Measured in a study.
From The stamp sawtooth model for SNUADC and BIS
About BIS against SNUADC, BIS EEG
Date 2026-10-19

Notes

The whole release (2,710 recordings with at least 5 good burst windows, model3). The pilot gave 33 ms median, p90 116 ms (16 and 67 ms with the case's own bursts) in 143 cases; the whole release includes recordings with no correction, with mode changes and with many gaps, which the pilot's cases mostly lacked.