BIS drift against the recorder PC¶
Drift of the BIS EEG stamps against the recorder PC, per case, from whole teeth.
| Value | 25 ppm (p10–p90 12 to 50; 5th–95th 7.3–51.1 ppm; census fit of 4,728 recordings median 21.1 ppm) |
| Status | measured: Measured in a study. |
| From | The stamp sawtooth model for SNUADC and BIS |
| About | BIS against the recorder PC |
| Date | 2026-10-19 |
By setup¶
Per recording setup (PC key: SNUADC · Solar8000 · Primus · BIS serial ports; see Recording setups). For a recording whose setup is known, use its row: the spread within a setup is far smaller than over all recordings.
| Setup | Value (ppm) | IQR | Cases | Note |
|---|---|---|---|---|
| COM4 · COM6 · COM5 · COM8 | 44.1 | 42.7 to 46.9 | 186 | |
| COM4 · COM12 · COM11 · COM14 | 50.5 | 49.7 to 51 | 178 | |
| COM19 · COM26 · COM27 · COM29 | 32.3 | 30.6 to 38.6 | 93 | |
| COM12 · COM4 · COM5 · COM7 | 21.5 | 20.5 to 22.6 | 75 | |
| COM4 · COM9 · COM10 · COM20 | 18.6 | 17.9 to 19.2 | 64 | |
| COM3 · COM4 · COM5 · COM7 | 25.1 | 23.3 to 32.2 | 60 | |
| COM13 · COM26 · COM27 · COM19 | 17.4 | 16.7 to 18.1 | 56 | |
| COM25 · COM26 · COM27 · COM29 | 30.9 | 29.2 to 32.4 | 56 | |
| COM4 · COM5 · COM6 · COM3 | 49.4 | 48.2 to 49.8 | 53 | |
| COM7 · COM4 · COM5 · COM9 | 18.9 | 17.8 to 19.7 | 51 | |
| COM18 · COM13 · COM14 · COM6 | 18.2 | 17.5 to 18.8 | 48 | |
| COM4 · COM5 · COM6 · COM8 | 54.7 | 52.6 to 55.2 | 44 | |
| COM4 · COM9 · COM10 · COM12 | 30.5 | 29.9 to 31.3 | 44 | |
| COM13 · COM14 · COM15 · COM19 | 18.4 | 17.8 to 19.4 | 39 | |
| COM4 · COM5 · COM6 · COM12 | 28.2 | 27.8 to 30 | 39 | |
| COM14 · COM10 · COM11 · COM18 | 25.6 | 24.8 to 26.5 | 38 | |
| COM45 · COM42 · COM43 · COM39 | 18.8 | 18.3 to 20.1 | 37 | |
| COM14 · COM4 · COM5 · COM7 | 26.3 | 25.7 to 26.8 | 36 | |
| COM15 · COM13 · COM14 · COM22 | 19.9 | 18.9 to 21.1 | 31 | |
| COM4 · COM11 · COM12 · COM8 | 23.6 | 22.8 to 24.5 | 30 | |
| COM4 · COM5 · COM6 · COM20 | 17.2 | 16.1 to 17.8 | 27 | |
| COM8 · COM42 · COM43 · COM39 | 18.7 | 18.1 to 19.8 | 26 | |
| COM4 · COM26 · COM27 · COM29 | 30.8 | 30.4 to 31.7 | 25 | |
| COM15 · COM25 · COM26 · COM30 | 24.4 | 23.9 to 24.8 | 23 | |
| COM15 · COM26 · COM27 · COM29 | 29.9 | 24.3 to 30.4 | 23 | |
| COM13 · COM4 · COM5 · COM23 | 19.3 | 18.5 to 20.7 | 20 | |
| COM8 · COM14 · COM15 · COM4 | 21 | 19.6 to 22.6 | 20 | |
| COM9 · COM15 · COM16 · COM20 | 26.4 | 26.1 to 27.2 | 20 | |
| COM13 · COM4 · COM5 · COM9 | 19.5 | 19.1 to 20.6 | 18 | |
| COM14 · COM21 · COM22 · COM18 | 24.7 | 24.3 to 25.1 | 18 | |
| COM31 · COM28 · COM29 · COM34 | 18.2 | 17.6 to 19.3 | 18 | |
| COM9 · COM10 · COM11 · COM16 | 17.8 | 17 to 18.6 | 18 | |
| COM3 · COM15 · COM16 · COM20 | 22.9 | 22.3 to 23.7 | 16 | |
| COM4 · COM11 · COM12 · COM3 | 49.1 | 48.1 to 50 | 14 | |
| COM15 · COM27 · COM26 · COM29 | 28.6 | 26.4 to 29.3 | 9 | |
| COM32 · COM28 · COM27 · COM30 | 42.6 | 42.5 to 42.6 | 9 | |
| COM4 · COM13 · COM14 · COM6 | 18.8 | 18.3 to 19.8 | 9 |
Notes¶
Full release, replacing the pilot value of 23 ppm (p10–p90 17–47). From the correction steps of all recordings (When each device's clock is corrected, in every recording, 19 October): median 24.8 ppm (5th–95th 7.3–51.1) in 1,939 recordings, identical to the clock-correction scan where both exist; the full census fit gives 4,728 recordings, median 21.1 ppm (5th–95th 2.1–52.8). Relative to the recorder PC.
Used in¶
- drift of BIS against the recorder PC