Skip to content

BIS drift against the recorder PC

Drift of the BIS EEG stamps against the recorder PC, per case, from whole teeth.

Value 25 ppm (p10–p90 12 to 50; 5th–95th 7.3–51.1 ppm; census fit of 4,728 recordings median 21.1 ppm)
Status measured: Measured in a study.
From The stamp sawtooth model for SNUADC and BIS
About BIS against the recorder PC
Date 2026-10-19

By setup

Per recording setup (PC key: SNUADC · Solar8000 · Primus · BIS serial ports; see Recording setups). For a recording whose setup is known, use its row: the spread within a setup is far smaller than over all recordings.

Setup Value (ppm) IQR Cases Note
COM4 · COM6 · COM5 · COM8 44.1 42.7 to 46.9 186
COM4 · COM12 · COM11 · COM14 50.5 49.7 to 51 178
COM19 · COM26 · COM27 · COM29 32.3 30.6 to 38.6 93
COM12 · COM4 · COM5 · COM7 21.5 20.5 to 22.6 75
COM4 · COM9 · COM10 · COM20 18.6 17.9 to 19.2 64
COM3 · COM4 · COM5 · COM7 25.1 23.3 to 32.2 60
COM13 · COM26 · COM27 · COM19 17.4 16.7 to 18.1 56
COM25 · COM26 · COM27 · COM29 30.9 29.2 to 32.4 56
COM4 · COM5 · COM6 · COM3 49.4 48.2 to 49.8 53
COM7 · COM4 · COM5 · COM9 18.9 17.8 to 19.7 51
COM18 · COM13 · COM14 · COM6 18.2 17.5 to 18.8 48
COM4 · COM5 · COM6 · COM8 54.7 52.6 to 55.2 44
COM4 · COM9 · COM10 · COM12 30.5 29.9 to 31.3 44
COM13 · COM14 · COM15 · COM19 18.4 17.8 to 19.4 39
COM4 · COM5 · COM6 · COM12 28.2 27.8 to 30 39
COM14 · COM10 · COM11 · COM18 25.6 24.8 to 26.5 38
COM45 · COM42 · COM43 · COM39 18.8 18.3 to 20.1 37
COM14 · COM4 · COM5 · COM7 26.3 25.7 to 26.8 36
COM15 · COM13 · COM14 · COM22 19.9 18.9 to 21.1 31
COM4 · COM11 · COM12 · COM8 23.6 22.8 to 24.5 30
COM4 · COM5 · COM6 · COM20 17.2 16.1 to 17.8 27
COM8 · COM42 · COM43 · COM39 18.7 18.1 to 19.8 26
COM4 · COM26 · COM27 · COM29 30.8 30.4 to 31.7 25
COM15 · COM25 · COM26 · COM30 24.4 23.9 to 24.8 23
COM15 · COM26 · COM27 · COM29 29.9 24.3 to 30.4 23
COM13 · COM4 · COM5 · COM23 19.3 18.5 to 20.7 20
COM8 · COM14 · COM15 · COM4 21 19.6 to 22.6 20
COM9 · COM15 · COM16 · COM20 26.4 26.1 to 27.2 20
COM13 · COM4 · COM5 · COM9 19.5 19.1 to 20.6 18
COM14 · COM21 · COM22 · COM18 24.7 24.3 to 25.1 18
COM31 · COM28 · COM29 · COM34 18.2 17.6 to 19.3 18
COM9 · COM10 · COM11 · COM16 17.8 17 to 18.6 18
COM3 · COM15 · COM16 · COM20 22.9 22.3 to 23.7 16
COM4 · COM11 · COM12 · COM3 49.1 48.1 to 50 14
COM15 · COM27 · COM26 · COM29 28.6 26.4 to 29.3 9
COM32 · COM28 · COM27 · COM30 42.6 42.5 to 42.6 9
COM4 · COM13 · COM14 · COM6 18.8 18.3 to 19.8 9

Notes

Full release, replacing the pilot value of 23 ppm (p10–p90 17–47). From the correction steps of all recordings (When each device's clock is corrected, in every recording, 19 October): median 24.8 ppm (5th–95th 7.3–51.1) in 1,939 recordings, identical to the clock-correction scan where both exist; the full census fit gives 4,728 recordings, median 21.1 ppm (5th–95th 2.1–52.8). Relative to the recorder PC.