The plethysmogram is built in 30 Hz frames

The SNUADC PLETH waveform comes from the oximeter in 33.3 ms frames, smoothed to 500 Hz. Its content changes once per frame, on the oximeter's own clock, about 0.4 s after the finger.

Status established: shown on many cases, with a control.
Clocks SNUADC sample clock
Devices Solar8000 monitor, SNUADC
Variables Plethysmogram (SNUADC)

What it means

Pulse arrival time from VitalDB's PLETH is about 0.4 s longer than physiology and carries a sawtooth of frame slips; raw-sample PPG fiducials lock to the frame grid.

What is done about it

physiosync removes the sawtooth; the planner shifts PLETH by its chain delay.

Evidence

Study Date Status Summary
Monitor clocks in the waveforms' spectral lines 2026-10-17 open Differences of the raw converter counts show lines from the monitor's analogue outputs: 240 Hz in arterial pressure and plethysmogram, and a 30 Hz family in the plethysmogram alone. Each measures a monitor clock against SNUADC.
The PPG hardware delay and the adopted pulse arrival time 2026-10-12 current VitalDB's raw ECG→PPG pulse arrival time is about 0.4 s longer than physiology. Splitting it into a hardware delay (one value for the release; One device delay for the PLETH across the release) and a physiological part needs an adopted value from studies that sampled ECG and PPG on one clock: about 270 ms for anaesthetised adults.
The PPG's 30 Hz frame clock 2026-10-10 current ECG and PPG fiducials resolve to one 2 ms sample, but PPG fiducials taken on raw samples lock to a hidden 33.3 ms frame grid. Scanning for it finds the oximeter's frame clock to about 1 ppm, a fixed fingerprint of each recording setup.
The PPG sawtooth and its correction 2026-10-10 current The ECG→PPG delay creeps at 18–52 ppm and jumps up by one 33.3 ms frame at each slip, every 11–31 minutes. A tooth-by-tooth model removes it; where the correction is confident it is essentially exact on injected test data.