PPG content against the oximeter frames (the sawtooth)

The PPG content runs 18–52 ppm against the oximeter's frames, fixed per setup. When a whole frame of lag has built up, one frame is repeated: the delay steps up by 33.3 ms, then creeps down again.

Clock Pulse oximeter frame clock
Measured against SNUADC sample clock
Modes all
Status measured
Drift 18–52 ppm by setup
Correction size 32.3 ms (IQR 30.4 to 34.8)

Correction

One repeated 33.3 ms frame whenever a frame of lag has built up (a slip). physiosync detects the slips in the beat-matched ECG→PPG delay and removes the sawtooth tooth by tooth.

Parameters

Parameter Value Status From
PPG sawtooth rate 18–52 ppm by setup measured The PPG sawtooth and its correction
PPG slip size 32.3 ms (IQR 30.4 to 34.8) measured The PPG sawtooth and its correction

Evidence

Study Date Status Summary
The PPG sawtooth and its correction 2026-10-10 current The ECG→PPG delay creeps at 18–52 ppm and jumps up by one 33.3 ms frame at each slip, every 11–31 minutes. A tooth-by-tooth model removes it; where the correction is confident it is essentially exact on injected test data.