The Solar8000 relays the Primus's ventilation numbers

Solar8000 ETCO₂, RR_CO2, VENT_PIP and VENT_RR change only every 6–8 s: they are Primus values passed on through the Solar8000's 2 s polling, not measured by the monitor.

Status established: shown on many cases, with a control.
Devices Solar8000 monitor, Primus anaesthesia machine

What it means

These tracks carry the Primus's timing plus the relay, and duplicate the Primus's own numerics.

Evidence

Study Date Status Summary
How the monitors compute their numbers 2026-10-12 current Recomputing each monitor numeric from its own waveform finds a short, simple algorithm for every one (a median, an exponential average or a time mean over 3–30 s) and a delay of 2–9 s behind the newest data it uses. No value is stamped before its data exist.