Capnogram (Primus)

Sidestream CO₂ from the Primus at 62.5 Hz; the sampling line adds at least 1 s.

Stub

Imported from the time-source table; the full page is still to be written.

Name in the dataset Primus/CO2
Type waveform
Device Primus anaesthesia machine
Stamped by Recorder PC clock
Spaced by Primus clock
Rate 62.5 Hz
Unit mmHg

Measurement space

Time side (after breath matching): the true time lies between -0.203 s and -0.055 s from the stamp: the delay range plus the interval one value describes.

Time

From UTC to the stored time stamp

Derived from this signal's timing table; a written chain is still to come.

The clock path first (UTC to the clock that stamps or spaces the samples), then the signal's own path to its stored time. Each bar: the time shift or error the step introduces (pale) and what is left after the atlas's correction (solid), on a log scale; dashed when estimated or assumed.
The reference every chain starts from.
sync assumed 3 ms – 900 ms · clock link, parameter
Windows time synchronisation to the hospital's time server every 8–24 h (assumed), drifting a few ppm in between. Not observable in the release: dates are shifted to 2100. No step of 0.5 s or more was found.
clock measured 600 ms – 5.1 s → 55 ms – 187 ms after correction
Delay or clock offset to the stamp; removed by the breath matching to SNUADC.
Stored time stamp
window measured 16 ms
The width of the interval one value or sample stands for.

Timing

Against the working reference. Delay: from the end of the interval a value describes to its stamp. Interval: how wide that interval is.

Form Delay Interval Clock domain Status Evidence
after breath matching (default) 55 ms – 187 ms 16 ms SNUADC stamp-model time (the working reference) measured leave-one-out error 55 ms median, p90 187 ms, a line per stretch between AWP pauses (1,067 recordings with CVP, tools/primus_sync_accuracy.py); 74 ms with one slope per recording
raw stamps 600 ms – 5.1 s 16 ms Primus clock measured offset p10-p90 0.6-5.1 s, median 1.8 s (1,061 recordings, examples/ppv-by-breath), incl. AWP-CVP physiology; drifts +104 ppm against the PC between AWP pauses

Combine it with other signals in the synchronisation planner.

Value

Its value side is still to be described.

Observations

Observation Status Summary
Eight recordings store whole messages twice established In 8 recordings 15–48% of the waveform records are byte-for-byte copies of the record before them, with the same timestamp: case 15, case 364, case 1946, case 3865, case 4857, case 5414, case 5595 and case 5619. SNUADC, Primus and BIS are affected equally, so the recorder, not a device, wrote them twice.
Primus waveform gains are wrong in every recording, for CO₂ or for AWP established Recordings store the Primus waveforms with one of three sets of scale factors. Checked against the Primus's own ETCO₂ and PIP numerics in all 6,388 recordings, CO₂ fits only 0.075 mmHg per code and AWP only 0.1 mbar per code with offset −20, so most recordings overstate CO₂ by a third, and the others overstate AWP by about 3% (950 recordings) or understate it about fourfold (26).
The Primus AWP stops for about 21 s every 7,230 s established In 5,502 recordings the Primus airway-pressure waveform has gaps of 19–23.5 s (whole messages of 0.48 s, mostly 21.12 s) that recur every 7,230 s, the Primus's correction interval; a clock correction follows about 21.6 s after. The CO₂ waveform carries on through 90% of them. The 21 s of AWP are lost, not merely re-stamped.
A heartbeat-locked trace in the CO₂ waveform that ignores the Primus clock open Averaged on the ECG's R peaks, the Primus CO₂ waveform shows a heartbeat-locked signal in 60% of 10-minute windows (30-recording pilot), tight to about 20 ms. Unlike the breath, its lag does not drift with the Primus clock: it stays fixed on the SNUADC timeline.

Studies

Study Date Status Summary
The heartbeat in the Primus airway signals 2026-10-19 current A pilot in 30 recordings for a heartbeat-locked route from the Primus to the SNUADC. Airway pressure shows the heartbeat in 9% of windows only. CO₂ shows a heartbeat-locked trace in 60% of windows, tight to 20 ms, but its lag does not follow the Primus drift, so it cannot time the Primus. Closed as a clock route.
Gaps and unbroken segments in every waveform 2026-10-19 draft Every gap and every unbroken segment in the 6,388 recordings, per waveform. The curves are not smooth: a cliff at about 18 s in every device marks the recorder's pauses, SNUADC segments stop at 5 hours, Primus segments at about 2.0 hours and BIS segments at about 2.3 hours (each device's correction cycle), and BIS's short gaps grow with the time since the last one (clock resynchronisations, not lost data).
When each device's clock is corrected, in every recording 2026-10-19 draft From the steps between records in all 6,388 recordings: the BIS is corrected every 8,192 s and the Primus every 7,230 s, on their own timers (the first correction comes at any time after the device starts); the SNUADC is reset when 0.49 s of drift has built up. With the full rerun of the correction fit, it gives each device's drift against the recorder PC in 4,728 BIS and 2,728 SNUADC recordings, and shows that the Primus's 20–23 s AWP pauses belong to its correction cycle. Every rate here is relative to the recorder PC, whose own clock drifts and is corrected.
The Primus stamp step at each airway-pressure pause 2026-10-19 current Measured directly at 2,491 pauses in 1,021 recordings: where CO₂ carries on through the AWP pause (74%), the Primus stamps barely step (median −1 ms, p10–p90 −164 to +187 ms, close to the no-pause controls); where CO₂ stops too (26%), they step by up to seconds. No step undoes the drift built up since the last pause.
Message sizes and the steps between records 2026-10-19 draft Every record in every waveform: SNUADC writes messages of 100 samples (0.2 s), Primus 30 (0.48 s), BIS 16 (0.125 s), CardioQ no fixed size. Backward steps outnumber gaps by about a hundred to one and come from a few recordings with duplicated or interleaved records. Primus gaps are whole lost messages; most SNUADC and BIS gaps are not.
What each break in the stamps is 2026-10-19 draft Every step between waveform records in the release, sorted by mechanism: written twice, a SNUADC clock reset, interleaved streams, a pause shared by every device, whole messages, less than a message, or a fraction of a message beyond whole ones. A continuity test with controls shows which classes lose data: shared pauses and most longer SNUADC gaps do; steps under one message, and SNUADC's lone one-message steps, mostly do not.
Value resolution of every signal 2026-10-19 draft How finely every waveform and numeric track in the release records its values. The SNUADC waveforms are 10-bit codes, though VitalRecorder's documentation describes the SNUADC as 12-bit; Primus waveforms use about 10 bits of 12; BIS EEG is a 16-bit grid of 0.05 µV. Numerics are 32-bit floats on decimal grids of 1, 0.1, 0.01 or 0.001, one per track; settings sit on coarser grids of 10 or 100.
How the monitors compute their numbers 2026-10-12 current Recomputing each monitor numeric from its own waveform finds a short, simple algorithm for every one (a median, an exponential average or a time mean over 3–30 s) and a delay of 2–9 s behind the newest data it uses. No value is stamped before its data exist.
The Primus ventilator against SNUADC, by breaths 2026-10-12 open Each breath shows in the Primus airway pressure and, through the chest, in SNUADC's central venous pressure. Tracking that lag puts the Primus on the SNUADC timeline to about 55 ms over the whole release, with a line per stretch between its AWP pauses (49 ms in the pilot). At scale it measures the Primus drifting a median +104 ppm against the recorder PC between its AWP pauses, with steps at the pauses that do not reset it.
Clock corrections in every file 2026-10-11 current A scan of all 6,388 files for the jumps where the recorder re-synchronises a device's stamps. Each device type has its own correction signature; only SNUADC and BIS correct clearly enough to measure their drift.

Tags: stub