Three lab analytes stop at a reporting limit

Lactate is capped at 15.0 mmol/L, aPTT at 400.0 s, and ESR does not go below 2 mm/h: the values at the limit stand for "this or beyond", not for the number itself.

Status likely: consistent, with a known gap.
Devices Clinical records
Variables Lactate, Activated partial thromboplastin time, Erythrocyte sedimentation rate

What it means

Treated as exact numbers, these values understate the most abnormal results: a lactate of "15" may be 20.

What is done about it

Treat them as censored (an open interval beyond the limit), as the registry does for ages above 89.

Detail

Analyte Limit Results at the limit Cases Most at any nearby value
Lactate 15.0 mmol/L (top) 317 28 16
aPTT 400.0 s (top) 226 87 8
ESR 2 mm/h (bottom) 89 72 32

Each limit holds many times more results than any neighbouring value, the pattern of an analyser's reportable range rather than of physiology.

Evidence

Study Date Status Summary
Heaping screen of all variables 2026-10-13 current Every hand-entered clinical value and every lab analyte was screened for mixed resolution and heaping, to decide which need a rounding model and which need only a registry entry.