Solar8000 heart rate comes one poll early or late

Heart rate reaches the recorder about 1.9 s or about 4.2 s after its window ends, one 2 s poll apart. The group belongs to the recording PC and stays fixed through a case.

Status established: shown on many cases, with a control.
Devices Solar8000 monitor, Vital Recorder
Variables Solar8000 heart rate

What it means

A single HR delay is wrong by about 2 s in a third of cases.

What is done about it

The hr_delay_group label gives each case's group; the PC gives a prior for uncertain cases.

Evidence

Study Date Status Summary
Two heart-rate delay groups, one poll apart 2026-10-13 current Solar8000 heart rate reaches the recorder either about 1.9 s or about 4.2 s after its window ends: one 2 s poll apart. The group belongs to the recording PC, stays fixed through a case, and on some PCs varies from case to case.