Value resolution of every signal

How finely every waveform and numeric track in the release records its values. The SNUADC waveforms are 10-bit codes, though VitalRecorder's documentation describes the SNUADC as 12-bit; Primus waveforms use about 10 bits of 12; BIS EEG is a 16-bit grid of 0.05 µV. Numerics are 32-bit floats on decimal grids of 1, 0.1, 0.01 or 0.001, one per track; settings sit on coarser grids of 10 or 100.

Date 2026-10-19
Status draft
Methods Rounding, heaping and value resolution
Devices SNUADC, Primus anaesthesia machine, BIS monitor, CardioQ, Solar8000 monitor, Infusion pumps, EV1000, Vigileo, Vigilance
Variables ECG lead II (SNUADC), Arterial pressure (SNUADC), Plethysmogram (SNUADC), Capnogram (Primus), Airway pressure (Primus), BIS EEG, BIS index, Solar8000 heart rate, SpO₂, Solar8000 ART and CVP numerics, Non-invasive blood pressure, Primus ETCO₂, PIP and RR, Infusion pump rates and volumes, EV1000 mean arterial pressure, Vigileo SV, CO and SVV, Vigilance continuous cardiac output

Background

A .vital file stores each track in a declared format with a gain and offset: measured value = offset + stored value × gain (VitalDB 2026). Waveforms from converters are stored as integer codes; numerics, and waveforms relayed from devices that send numbers, as 32-bit floats. The format says how many bits are available, not how many the device used: that has to be read from the values.

Data

All 6,388 recordings: every sample of every waveform track (a full histogram of stored values per track) and every record of every numeric track (each distinct value with its count).

Method

  • Waveforms. One bin per stored code over the whole release. The resolution is the spacing of the codes; the bits to store the observed span are the whole number of bits that hold every code between its ends. The range is taken between the 10⁻⁷ and 1 − 10⁻⁷ quantiles of the samples, widened to any code within 64 steps of the body, so that a handful of corrupt samples cannot set it. The spacing is also read in 8 bands of equal sample count, which shows a scale that is not uniform.
  • Numerics. The share of records on each decimal grid (100, 10, 5, 1, 0.5, … 0.001), allowing a few units of float32 rounding (a stored 0.1 reads 0.099999…). The resolution is the coarsest grid holding 99% of records; the share on the next coarser grid shows heaping or a mixed resolution (by chance, about 1/2, 1/5 or 1/10).

Results

Waveforms. Two quantities are kept apart. The observed span is what the data show: the codes between the lowest and highest values the signal reaches (on the step it uses), and the whole number of bits needed to store that span. The converter resolution is an estimate of the device: it can be no less than the observed span, and it is stated only with its evidence (all codes of a power-of-two range used, a documented value, or rare values at the edge of a larger range).

Track Stored as Step Observed span (codes used) Bits to store the span Converter resolution (estimate) Evidence
SNUADC ECG II, V5 uint16 0.00987 mV 0–1023 (all 1,024) 10 10 bits every code of 0–1023 used, none above; documented as 12-bit
SNUADC ART, CVP, FEM uint16 0.987 mmHg 0–983 (848) 10 10 bits same converter as ECG
SNUADC PLETH uint16 0.395 0–1023 (852) 10 10 bits same converter as ECG
Primus CO₂ uint16 0.1 mmHg as stored (0.075 fits) 0–1564 (1,530); 8 samples to 4007 11 probably 12 bits rare values up to 4046 in CO₂ and AWP, inside 0–4095
Primus AWP uint16 0.1 mbar 0–1032 (1,010); 22 samples to 4046 11 probably 12 bits as CO₂
BIS EEG1, EEG2 float32 0.05 µV −1476.70 to +1800.05 (all 65,536); 400 corrupt samples 16 16 bits every code of a 2¹⁶ range used, none outside it
CardioQ FLOW float32 1, 2 or 4 (segmented) 0–1016 (509 values) 9 as codes (10 as values) unknown a 9-bit segmented code (CardioQ FLOW uses three scales (a segmented code))
CardioQ ABP float32 1 mmHg −100 to 414 (497); 2 corrupt samples 10 unknown, at least 10 bits whole mmHg, a display resolution rather than a converter
  • SNUADC is 10-bit in the data, 12-bit in the documentation. VitalRecorder's device list describes the SNUADC as an 8-channel 12-bit, 500 Hz converter (VitalDB 2026). Every SNUADC channel in the release stores codes 0–1023 and nothing above, so the stored values have 10 bits whatever the converter delivers (SNUADC values are 10-bit codes, though the device is documented as 12-bit).
  • One volt scale behind every SNUADC channel. The SNUADC is configured with one gain per channel, from volts to the physical unit (VitalDB 2026). The stored gains are that gain times one fixed step of 0.0098746 V per code: ECG gain 1, ART, CVP and FEM gain 100, PLETH gain 40, and the offsets are −4.956 V times the same gain. The converter therefore spans −4.956 to +5.156 V in 1,024 codes.
  • Primus waveforms are stored in 16 bits but use about 10; their few samples near code 4,000 fit a 12-bit device range (0–4095). Their stored gains are not all right: against the Primus's own numerics, CO₂ fits only 0.075 mmHg per code and AWP only 0.1 mbar per code, so most recordings overstate CO₂ by a third (Primus waveform gains are wrong in every recording, for CO₂ or for AWP).
  • The earlier summary counted 54.9 bits for BIS EEG and 16.5 for CardioQ ABP, set by a few corrupt samples; with the robust range the observed spans need 16 and 10 bits.

Numerics (the full table, 150 tracks, is in the work folder: numeric_resolution.csv)

Resolution Tracks
100 Primus fresh-gas flow setting; N2O flow and N2O fractions (almost always 0)
10 FiO₂ and tidal volume settings (Primus and Solar8000)
1 Solar8000 HR, pulse rate, SpO₂, ART, CVP, PA, NIBP, ETCO₂, FiO₂, ventilator pressures and RR; Primus ETCO₂, FiO₂, pressures, RR, TV, gas flows; EV1000 / Vigileo SV, SVI, SVR; Vigilance volumes and SvO₂; INVOS
0.1 Temperature; ST segments; minute volume; MAC and agent concentrations (kPa); compliance; inspiratory time; BIS index, SQI and SR; CO and CI (EV1000, Vigileo, Vigilance, CardioQ); EV1000 SVV; Orchestra target concentrations (CT) and the rates of most drugs
0.01 BIS SEF, EMG and total power; Primus TV setting (L, so 10 mL); Solar8000 gas 2 (94% on 0.1)
0.001 Orchestra volumes, plasma and effect-site concentrations (CP, CE), and the propofol and remifentanil rates
  • Each track has one resolution throughout; none mixes grids except Solar8000 gas 2 (94% on 0.1, the rest on 0.01) and Solar8000 FiO₂ / FeO₂ (0.08% off whole percent).
  • The same quantity from two devices agrees: Primus and Solar8000 minute volume (0.1 L/min), TV (1 mL), and the TV setting (0.01 L on the Primus, 10 mL on the Solar8000).
  • Corrupt values appear in a few tracks: BIS SQI to −2.1 × 10⁸, BIS index to −1,333, Orchestra concentrations and volumes to 10⁶, Primus mean airway pressure 4,000. They will be listed with the leads-off and no-signal codes.

Discussion

Recorded values are as fine as the stored codes or the decimal grid, not as fine as the storage format. For waveforms, the step is the quantisation of the measurement (uniform within ±½ step, 1 SD = step/√12; truncated rather than rounded in the SNUADC, SNUADC values are truncated to the count below). For numerics, the grid is the device's display resolution; whether the device rounded or truncated to it is known only for the Solar8000 heart rate (truncated) and respiratory rate (rounded) (Solar8000 heart rate is truncated, not rounded).

Observations it supports

Observation Status Summary
Primus waveform gains are wrong in every recording, for CO₂ or for AWP established Recordings store the Primus waveforms with one of three sets of scale factors. Checked against the Primus's own ETCO₂ and PIP numerics in all 6,388 recordings, CO₂ fits only 0.075 mmHg per code and AWP only 0.1 mbar per code with offset −20, so most recordings overstate CO₂ by a third, and the others overstate AWP by about 3% (950 recordings) or understate it about fourfold (26).
CardioQ FLOW uses three scales (a segmented code) likely CardioQ FLOW values step by 1 from 0 to 255, by 2 from 256 to 510 and by 4 from 512 to 1016: 509 values, a 9-bit code whose step doubles in each upper segment, as in µ-law and A-law companding or a small floating-point format (an exponent and a 7-bit mantissa). Below 256, even values outnumber odd ones by 40 to 230 times.
SNUADC values are 10-bit codes, though the device is documented as 12-bit open Every SNUADC channel in the release stores codes 0–1023 and nothing above: 10 bits, a step of 0.0098746 V. The VitalRecorder device list describes the SNUADC as a 12-bit converter.

Limitations

  • The documentation describes the current VitalRecorder; the release was recorded with earlier versions and SNUADC hardware whose resolution may differ from the documented one.
  • The decimal-grid test cannot tell whether a float numeric started as an integer count on a finer scale (a device's own converter); that is a separate question.

Open questions

  • Why 10 bits from a 12-bit converter? (weakly supported hypothesis). Either the SNUADC used in the release was a 10-bit design, or a 12-bit converter's readings are reduced to 10 bits before they are sent (which the observed truncation would fit). The SNUADC firmware or the VitalDB team could settle it.
  • Rounding or truncation in each numeric (next step). Against the Primus's own waveforms (with the corrected gains), all 6,342 recordings favour rounding for ETCO₂ (where round and floor differ, 77% of numerics match round, 18% floor), PIP (61% / 32%) and mean airway pressure (62% / 28%). The match is not complete because the window does not reproduce the device's own computation, so the result is "likely". The Solar8000 ART numerics cannot be tested this way: the SNUADC copy of the arterial pressure differs from them by more than one step.
  • Integer codes behind float signals (next step). Signals sent as floats (BIS EEG, CardioQ, the numerics) may all start as integers from a converter of n bits, scaled once. Recover the integer and the scale for each (BIS EEG already fits 16 bits of 0.05 µV; CardioQ FLOW looks companded).

Code

References

  • VitalDB. VitalRecorder supported devices. VitalDB documentation. Accessed 19 October 2026. ↗
  • VitalDB. VitalRecorder configuration file (vr.conf) guide. VitalDB documentation. Accessed 19 October 2026. ↗
  • VitalDB. Vital file format specification (format version 3). VitalDB documentation. Accessed 19 October 2026. ↗