Clocks¶
Anything that keeps time: a crystal, a computer clock, a frame clock, a pacemaker.
| BIS clock | The BIS monitor's own clock. |
| CardioQ clock | The CardioQ's own clock; the recorder re-anchors its stamps about every 5 minutes. |
| Hospital records clock | The clocks of the anaesthesia record and the laboratory system, and the rounding of what was entered. |
| Monitor analogue output clock | A line at 240 Hz in the arterial pressure and plethysmogram (weakly in the ECG) runs −25 to −29 ppm against the SNUADC sample clock: most likely the patient monitor's output stage. |
| Pacemaker clock | The crystal of an implanted pacemaker. At a fixed pacing rate it beats at one constant interval that no device in the recording chain can touch, which makes it an independent reference. |
| Primus clock | The Primus anaesthesia machine's own clock. |
| Pulse oximeter frame clock | The oximeter module plays the plethysmogram out in 33.3 ms (30 Hz) frames on its own clock, fixed per recording setup at +7 to +92 ppm against the SNUADC sample clock (medians over 42 PC keys). |
| Recorder PC clock | The Windows clock of the PC running Vital Recorder. It stamps every record, and every other clock is linked to it, directly or through others. |
| SNUADC sample clock | The sample clock of the SNUADC acquisition board, which digitises ECG, arterial pressure, plethysmogram and CVP at 500 Hz. The working reference for every waveform alignment. |
| Solar8000 clock | The Solar8000 monitor's internal timing of its numeric windows. |
| UTC | Coordinated Universal Time, the reference every chain of clocks in the atlas starts from. |