The cardiac artefact in the BIS EEG

The BIS EEG carries a weak electrical trace of every heartbeat. Too weak for beat-by-beat matching, it shows clearly when averaged on the ECG's R peaks, and gives a second SNUADC–BIS timing reference that agrees with the bursts to about 16 ms.

Date 2026-10-12
Status current
Methods One event, two clocks
Techniques Null controls
Clocks BIS clock, SNUADC sample clock
Devices BIS monitor, SNUADC
Variables BIS EEG, ECG lead II (SNUADC)

Background

Heartbeat interval matching aligns EEG with ECG when the EEG carries the heartbeat (Goodwin et al. 2023). A first attempt found no usable heartbeat in the BIS EEG; it searched −0.3 to +0.5 s around each R peak on raw stamps, while the same event is stamped about 0.8 s earlier in BIS.

Data

30 benchmark cases with BIS (a first attempt on 28), three 10-minute windows per channel each.

Method

Both devices on stamp-model time; R-triggered averages of the EEG from −2 to +1 s, in three 10-minute windows per channel per case; the peak lag compared with the burst-measured offset.

Results

  • A clear peak (peak-to-noise ≥ 8) in 44 of 180 windows, in 13 of 30 cases; weaker peaks at the same lag in many more. Median amplitude 4.3 µV.
  • Its lag matches the burst offset to a median of −16 ms (p10–p90 −31 to +9 ms): electrical, near-simultaneous with the R wave.
  • Single beats are too weak: single-beat SNR 0.4 median, so beat-by-beat interval matching stays impractical.

Discussion

The averaged artefact is a second, independent SNUADC–BIS reference, including in cases with no bursts. It supersedes the first attempt's conclusion that the BIS EEG carries no heartbeat.

Observations it supports

Observation Status Summary
The BIS EEG carries an averaged cardiac artefact likely Averaged on the ECG's R peaks, the BIS EEG shows the heartbeat (median 4.3 µV), near-simultaneous with the R wave. Single beats are too weak to match one by one.

Open questions

  • Why the artefact shows in only some cases (unexplained phenomenon). A clear averaged artefact appears in 13 of 30 cases. What decides whether it is visible (electrode placement, signal quality, filtering) is not known.

Code

References

  • Goodwin AJ, Dixon W, Mazwi M, Hahn CD, Meir T, Goodfellow SD, Kazazian V, Greer RW, McEwan A, Laussen PC, et al. The truth Hertz – synchronization of electroencephalogram signals with physiological waveforms recorded in an intensive care unit. Physiol Meas. 2023;44(8):085002. ↗