Primus anaesthesia machine

Dräger Primus: airway pressure and CO₂ waveforms at 62.5 Hz and ventilation numerics.

Stub

Imported from the time-source table; the full page is still to be written.

Type ventilator
Clocks Primus clock

Device documentation

Variables

Variable Type Stamped by Spaced by
Airway pressure (Primus) waveform Recorder PC clock Primus clock
Capnogram (Primus) waveform Recorder PC clock Primus clock
Primus ETCO₂, PIP and RR numeric Recorder PC clock —

Scale factors and resolution

Waveform scale factors. Physical value = offset + stored code × gain, one row per stored configuration. A factor shown wrong has its correction in Corrected to.

Track Unit Stored as Rate Gain / offset Recordings Corrected to
AWP hPa uint16 62.5 Hz 0.1 / -20 5,384 —
AWP hPa uint16 62.5 Hz 0.1064 / -22 950 0.1 / -20: PIP check (evidence)
AWP hPa uint16 62.5 Hz 0.01 / 0 26 0.1 / -20: PIP check (evidence)
CO2 mmHg uint16 62.5 Hz 0.1 / 0 5,412 0.075 / 0: ETCO₂ check (evidence)
CO2 mmHg uint16 62.5 Hz 0.075 / 0 950 —

Waveform resolution. Observed span is what the data reach, and the whole bits needed to store it; converter is our estimate of the device's resolution (never less than the span), with its evidence. The step is in the physical unit, with the corrected gain where the stored one is wrong.

Track Step Observed span (codes used) Bits to store span Converter (estimate) Evidence
AWP 0.1 codes 0–1032 (1,010 used); 22 corrupt samples outside 11 probably 12 bits rare values up to code 4046 in CO₂ and AWP, inside 0–4095
CO2 0.075 codes 0–1564 (1,530 used); 8 corrupt samples outside 11 probably 12 bits rare values up to code 4046 in CO₂ and AWP, inside 0–4095

Numerics. Stored as floats with gain 1. Grid is the coarsest decimal step holding 99% of records; the share on the next coarser grid shows heaping (by chance about 1/2, 1/5 or 1/10).

Track Unit Stored as Interval Recordings Grid Share one step coarser Rounding Observed range Note
PAMB_MBAR mbar float32 7.01 s 6,361 1 0.2 — 971 to 1030 —
SET_AGE — float32 7.01 s 6,361 1 0.21 — 0 to 94 —
VENT_LEAK mL/min float32 7.01 s 6,349 1 0.22 — 5 to 394 —
ETCO2 mmHg float32 7.02 s 6,342 1 0.22 round, likely (Value resolution of every signal) 0 to 86 —
INCO2 mmHg float32 7.02 s 6,342 1 0.23 — 0 to 48 —
MAWP_MBAR mbar float32 7.02 s 6,342 1 0.19 round, likely (Value resolution of every signal) -6 to 33 corrupt values to 4e+03
FEN2O % float32 7.02 s 6,339 100 — — 0 to 55 —
FIN2O % float32 7.02 s 6,339 100 — — 0 to 66 —
MAC — float32 7.02 s 6,339 0.1 0.7 — 0 to 4 —
FEO2 % float32 7.02 s 6,337 1 0.24 — 12 to 100 —
FIO2 % float32 7.02 s 6,337 1 0.24 — 16 to 100 —
RR_CO2 /min float32 7.02 s 6,333 1 0.21 round (How the monitors compute their numbers) 0 to 55 —
MV L float32 7.02 s 6,085 0.1 0.2 — 0 to 20 —
SET_FIO2 % float32 7.02 s 6,047 10 0.14 — 21 to 100 —
SET_FRESH_FLOW mL/min float32 7.03 s 6,047 100 — — 200 to 18000 —
PIP_MBAR mbar float32 7.02 s 6,025 1 0.21 round, likely (Value resolution of every signal) -2 to 57 —
TV mL float32 7.02 s 6,025 1 0.2 — 7 to 2450 —
COMPLIANCE mL / mbar float32 7.02 s 6,020 0.1 0.2 — 0.5 to 97.3 —
PEEP_MBAR mbar float32 7.02 s 6,003 1 0.84 — -2 to 19 —
PPLAT_MBAR mbar float32 7.02 s 6,003 1 0.21 — 0 to 49 —
SET_INSP_TM sec float32 7.04 s 5,982 0.1 0.21 — 0.6 to 3.5 —
SET_INTER_PEEP mbar float32 7.04 s 5,982 1 0.96 — 0 to 13 —
SET_RR_IPPV /min float32 7.04 s 5,982 1 0.21 — 7 to 34 —
SET_INSP_PAUSE % float32 7.04 s 5,973 10 0.005 — 0 to 20 —
SET_PIP mbar float32 7.04 s 5,973 1 0.81 — 14 to 61 —
SET_TV_L L float32 7.04 s 5,973 0.01 0.41 — 0.03 to 0.82 —
FLOW_AIR mL / min float32 7.01 s 5,645 1 0.41 — 0 to 9999 —
FLOW_O2 mL / min float32 7.01 s 5,644 1 0.34 — 0 to 9999 —
FLOW_N2O mL/min float32 7.01 s 5,643 100 — — 0 to 2926 —
EXP_SEVO kPa float32 7.02 s 3,733 0.1 0.57 — 0 to 7.1 —
INSP_SEVO kPa float32 7.02 s 3,733 0.1 0.6 — 0 to 7.3 —
EXP_DES kPa float32 7.02 s 2,156 0.1 0.68 — 0 to 11 —
INSP_DES kPa float32 7.02 s 2,156 0.1 0.7 — 0 to 12.8 —
SET_INSP_PRES mbar float32 7.08 s 431 1 0.3 — 5 to 60 —
SET_FLOW_TRIG L/min float32 7.14 s 247 0.1 0.98 — 0.3 to 15 —

Observations

Observation Status Summary
Eight recordings store whole messages twice established In 8 recordings 15–48% of the waveform records are byte-for-byte copies of the record before them, with the same timestamp: case 15, case 364, case 1946, case 3865, case 4857, case 5414, case 5595 and case 5619. SNUADC, Primus and BIS are affected equally, so the recorder, not a device, wrote them twice.
Monitor numerics describe the past, never the future established Every monitor numeric describes a window of 3–30 s or more that ended 1–9 s before its stamp. None is stamped before its data exist, so the stamps already respect the availability rule.
Numeric delays are fixed within a case but differ between cases established A numeric's delay does not drift through a recording, but 40–70% of its variance lies between cases. One delay per case describes it; one delay for the whole dataset is wrong by up to about 2 s.
Primus waveform gains are wrong in every recording, for CO₂ or for AWP established Recordings store the Primus waveforms with one of three sets of scale factors. Checked against the Primus's own ETCO₂ and PIP numerics in all 6,388 recordings, CO₂ fits only 0.075 mmHg per code and AWP only 0.1 mbar per code with offset −20, so most recordings overstate CO₂ by a third, and the others overstate AWP by about 3% (950 recordings) or understate it about fourfold (26).
The Primus AWP stops for about 21 s every 7,230 s established In 5,502 recordings the Primus airway-pressure waveform has gaps of 19–23.5 s (whole messages of 0.48 s, mostly 21.12 s) that recur every 7,230 s, the Primus's correction interval; a clock correction follows about 21.6 s after. The CO₂ waveform carries on through 90% of them. The 21 s of AWP are lost, not merely re-stamped.
The Solar8000 relays the Primus's ventilation numbers established Solar8000 ETCO₂, RR_CO2, VENT_PIP and VENT_RR change only every 6–8 s: they are Primus values passed on through the Solar8000's 2 s polling, not measured by the monitor.
A heartbeat-locked trace in the CO₂ waveform that ignores the Primus clock open Averaged on the ECG's R peaks, the Primus CO₂ waveform shows a heartbeat-locked signal in 60% of 10-minute windows (30-recording pilot), tight to about 20 ms. Unlike the breath, its lag does not drift with the Primus clock: it stays fixed on the SNUADC timeline.

Studies

Study Date Status Summary
The heartbeat in the Primus airway signals 2026-10-19 current A pilot in 30 recordings for a heartbeat-locked route from the Primus to the SNUADC. Airway pressure shows the heartbeat in 9% of windows only. CO₂ shows a heartbeat-locked trace in 60% of windows, tight to 20 ms, but its lag does not follow the Primus drift, so it cannot time the Primus. Closed as a clock route.
Gaps and unbroken segments in every waveform 2026-10-19 draft Every gap and every unbroken segment in the 6,388 recordings, per waveform. The curves are not smooth: a cliff at about 18 s in every device marks the recorder's pauses, SNUADC segments stop at 5 hours, Primus segments at about 2.0 hours and BIS segments at about 2.3 hours (each device's correction cycle), and BIS's short gaps grow with the time since the last one (clock resynchronisations, not lost data).
When each device's clock is corrected, in every recording 2026-10-19 draft From the steps between records in all 6,388 recordings: the BIS is corrected every 8,192 s and the Primus every 7,230 s, on their own timers (the first correction comes at any time after the device starts); the SNUADC is reset when 0.49 s of drift has built up. With the full rerun of the correction fit, it gives each device's drift against the recorder PC in 4,728 BIS and 2,728 SNUADC recordings, and shows that the Primus's 20–23 s AWP pauses belong to its correction cycle. Every rate here is relative to the recorder PC, whose own clock drifts and is corrected.
The Primus stamp step at each airway-pressure pause 2026-10-19 current Measured directly at 2,491 pauses in 1,021 recordings: where CO₂ carries on through the AWP pause (74%), the Primus stamps barely step (median −1 ms, p10–p90 −164 to +187 ms, close to the no-pause controls); where CO₂ stops too (26%), they step by up to seconds. No step undoes the drift built up since the last pause.
Message sizes and the steps between records 2026-10-19 draft Every record in every waveform: SNUADC writes messages of 100 samples (0.2 s), Primus 30 (0.48 s), BIS 16 (0.125 s), CardioQ no fixed size. Backward steps outnumber gaps by about a hundred to one and come from a few recordings with duplicated or interleaved records. Primus gaps are whole lost messages; most SNUADC and BIS gaps are not.
What each break in the stamps is 2026-10-19 draft Every step between waveform records in the release, sorted by mechanism: written twice, a SNUADC clock reset, interleaved streams, a pause shared by every device, whole messages, less than a message, or a fraction of a message beyond whole ones. A continuity test with controls shows which classes lose data: shared pauses and most longer SNUADC gaps do; steps under one message, and SNUADC's lone one-message steps, mostly do not.
Value resolution of every signal 2026-10-19 draft How finely every waveform and numeric track in the release records its values. The SNUADC waveforms are 10-bit codes, though VitalRecorder's documentation describes the SNUADC as 12-bit; Primus waveforms use about 10 bits of 12; BIS EEG is a 16-bit grid of 0.05 µV. Numerics are 32-bit floats on decimal grids of 1, 0.1, 0.01 or 0.001, one per track; settings sit on coarser grids of 10 or 100.
Identifying the recording setup of every case 2026-10-18 open Recognising which equipment made each recording, from its serial ports and, where those fall short, from clock fingerprints, so that per-setup values can be used for every case.
How often each device is corrected 2026-10-17 current The devices re-synchronise in different ways: SNUADC when its drift reaches about half a second, BIS on a fixed 8,192 s timer, and the Primus by small steps about every two hours that do not keep up with its drift (+104 ppm, measured later by breath matching).
Recorder PC clock steps and recording pauses 2026-10-17 current A step of the recorder PC clock would move every device's stamps at once. Searching all 6,388 files for such simultaneous jumps found none of 0.5 s or more; what moves everything together is recording pauses.
Closing loops in the timing graph 2026-10-13 current Where two independent routes reach the same clock, the offsets around the loop must sum to zero. Three measured loops close within their noise; the BIS loop closes up to a constant latency; the proposed EMR loop is not a loop at all.
Anaesthesia record times are rounded up 2026-10-12 current Anaesthesia start and end entered on a 5-minute mark sit 90–290 s later, against device markers, than entries typed to the minute. Rounding up to the next 5 minutes fits every one of 11 markers better than rounding to the nearest or down.
Numeric delays are fixed within a case, different between cases 2026-10-12 current With each numeric's algorithm held fixed, delays were fitted for every case and every 30-minute block. They do not drift within a case, but 40–70% of their variance lies between cases, so a single population delay is wrong by up to 2 s for HR and 1 s for the others. Each case gets its own delay label.
How the monitors compute their numbers 2026-10-12 current Recomputing each monitor numeric from its own waveform finds a short, simple algorithm for every one (a median, an exponential average or a time mean over 3–30 s) and a delay of 2–9 s behind the newest data it uses. No value is stamped before its data exist.
The Primus ventilator against SNUADC, by breaths 2026-10-12 open Each breath shows in the Primus airway pressure and, through the chest, in SNUADC's central venous pressure. Tracking that lag puts the Primus on the SNUADC timeline to about 55 ms over the whole release, with a line per stretch between its AWP pauses (49 ms in the pilot). At scale it measures the Primus drifting a median +104 ppm against the recorder PC between its AWP pauses, with steps at the pauses that do not reset it.
Clock corrections in every file 2026-10-11 current A scan of all 6,388 files for the jumps where the recorder re-synchronises a device's stamps. Each device type has its own correction signature; only SNUADC and BIS correct clearly enough to measure their drift.

Tags: stub