The SNUADC firmware truncates each calibrated value instead of rounding it, so every value sits half a count low on average (about 0.5 mmHg for arterial pressure). Values clamp at counts 0 and 1023.

Status likely: consistent, with a known gap.
Devices SNUADC
Variables Arterial pressure (SNUADC), ECG lead II (SNUADC), Plethysmogram (SNUADC)

What it means

A small constant bias, and clipping at the monitor's ±4 V output limits (arterial pressure near 400 mmHg).

Evidence

Study Date Status Summary
The SNUADC firmware 2026-10-13 current The public SNUADC firmware explains the board's timing: it sends no time stamps, samples on its 16 MHz crystal, loses time rather than bunching samples after a stall, reads its channels one after another, and truncates its values. A predicted fault at the microsecond counter's wrap was not found in the data.