Arterial pressure and aortic flow waveforms from the CardioQ oesophageal Doppler monitor.

Stub

Imported from the time-source table; the full page is still to be written.

Name in the dataset CardioQ/ABP
Type waveform
Device CardioQ
Stamped by Recorder PC clock
Spaced by CardioQ clock
Rate 180 Hz

Measurement space

Time side (after beat matching): the true time lies between -0.3056 s and -0.1 s from the stamp: the delay range plus the interval one value describes.

Time

From UTC to the stored time stamp

Derived from this signal's timing table; a written chain is still to come.

The clock path first (UTC to the clock that stamps or spaces the samples), then the signal's own path to its stored time. Each bar: the time shift or error the step introduces (pale) and what is left after the atlas's correction (solid), on a log scale; dashed when estimated or assumed.
The reference every chain starts from.
sync assumed 3 ms – 900 ms · clock link, parameter
Windows time synchronisation to the hospital's time server every 8–24 h (assumed), drifting a few ppm in between. Not observable in the release: dates are shifted to 2100. No step of 0.5 s or more was found.
clock measured 400 ms – 4.4 s → 100 ms – 300 ms after correction
Delay or clock offset to the stamp; removed by the beat matching to SNUADC ART.
Stored time stamp
window measured 5.6 ms
The width of the interval one value or sample stands for.

Timing

Against the working reference. Delay: from the end of the interval a value describes to its stamp. Interval: how wide that interval is.

Form Delay Interval Clock domain Status Evidence
after beat matching (default) 100 ms – 300 ms 5.6 ms SNUADC stamp-model time (the working reference) measured per stretch; matched-beat MAD 0.24 s, R-R correlation 0.96 (24 cases)
raw stamps 400 ms – 4.4 s 5.6 ms CardioQ clock measured lag vs SNUADC ART p5-p95 0.4-4.4 s; re-anchors ~40 times per case

Combine it with other signals in the synchronisation planner.

Value

Its value side is still to be described.

Studies

Study Date Status Summary
Closing loops in the timing graph 2026-10-13 current Where two independent routes reach the same clock, the offsets around the loop must sum to zero. Three measured loops close within their noise; the BIS loop closes up to a constant latency; the proposed EMR loop is not a loop at all.
Rare devices against SNUADC 2026-10-13 current CardioQ, EV1000, Vigileo and Vigilance each get a link to the SNUADC timeline: beat matching for the CardioQ, mean arterial pressure alignment for the EV1000 and Vigileo, and a long-window fit for Vigilance's averaged heart rate.