Monitor numerics describe the past, never the future

Every monitor numeric describes a window of 3–30 s or more that ended 1–9 s before its stamp. None is stamped before its data exist, so the stamps already respect the availability rule.

Status established: shown on many cases, with a control.
Devices Solar8000 monitor, Primus anaesthesia machine, BIS monitor
Variables Solar8000 heart rate, Solar8000 ART and CVP numerics, Primus ETCO₂, PIP and RR, BIS index, SpO₂, Solar8000 pulse rate (PLETH_HR)

What it means

Aligning a numeric with the waveform it came from means shifting it back by its delay and spreading it over its window. Left at its stamp, it is safe for real-time models but late for physiology.

What is done about it

The TimeNet import labels each numeric track with its typical delay and window (numeric_lag), and per-case labels carry each case's own delay. Stored times are never moved.

Evidence

Study Date Status Summary
How the BIS index trails its EEG 2026-10-13 current Modelling the BIS index as a lagged mix of EEG features shows a trailing average starting about 3 s after its EEG window ends: about 16 s long in half the cases (the 15 s smoothing setting) and 32–61 s in most others. Half the response weight is older than 30 s.
How the monitors compute their numbers 2026-10-12 current Recomputing each monitor numeric from its own waveform finds a short, simple algorithm for every one (a median, an exponential average or a time mean over 3–30 s) and a delay of 2–9 s behind the newest data it uses. No value is stamped before its data exist.
Numeric delays are fixed within a case, different between cases 2026-10-12 current With each numeric's algorithm held fixed, delays were fitted for every case and every 30-minute block. They do not drift within a case, but 40–70% of their variance lies between cases, so a single population delay is wrong by up to 2 s for HR and 1 s for the others. Each case gets its own delay label.