Many "lost pulse" events in the arterial pressure are an NIBP cuff inflating on the same arm: the pulse fades and pressure decays to 35–50 mmHg, then returns as the cuff cycle ends.

Status established: shown on many cases, with a control.
Devices Solar8000 monitor, SNUADC
Variables Arterial pressure (SNUADC), Non-invasive blood pressure

What it means

Cuff inflations disturb arterial timing as much as line accesses do.

What is done about it

lineaccess labels them occlusions (art_line:occlusion), separately from accesses.

Evidence

Study Date Status Summary
Arterial line accesses in the pressure waveform 2026-10-11 current Flushes, zeroing and blood draws leave clear signatures in the arterial pressure. A detector finds them in the waveform alone (about one access per hour) and matches 77% of lab draws to an access within 15 minutes, against 26% for random times.