Skip to content

The BIS EEG carries an averaged cardiac artefact

Averaged on the ECG's R peaks, the BIS EEG shows the heartbeat (median 4.3 µV), near-simultaneous with the R wave. Single beats are too weak to match one by one.

Status likely: consistent, with a known gap.
Devices BIS monitor, SNUADC
Variables BIS EEG, ECG lead II (SNUADC)
Supersedes The BIS EEG carries no heartbeat

What it means

A second SNUADC–BIS timing reference for cases without bursts. Shown on 30 cases.

Evidence

Study Date Status Summary