Oximeter frames against SNUADC

The oximeter's 33.3 ms frame period is fixed per setup against the SNUADC sample clock, +7 to +92 ppm across 42 PC keys, and keeps its phase through the sawtooth slips.

Clock Pulse oximeter frame clock
Measured against SNUADC sample clock
Modes all
Status measured
Drift +7 to +92 ppm by setup (medians over 42 PC keys); within 3 ppm of its setup's median in 89% of recordings

Parameters

Parameter Value Status From
Oximeter frame clock against SNUADC +7 to +92 ppm by setup (medians over 42 PC keys); within 3 ppm of its setup's median in 89% of recordings measured The PPG's 30 Hz frame clock

Evidence

Study Date Status Summary
The PPG's 30 Hz frame clock 2026-10-10 current ECG and PPG fiducials resolve to one 2 ms sample, but PPG fiducials taken on raw samples lock to a hidden 33.3 ms frame grid. Scanning for it finds the oximeter's frame clock to about 1 ppm, a fixed fingerprint of each recording setup.