Solar8000 ST records are written twice¶
Every Solar8000 ST_II record is duplicated at the same stamp (0–20 ms apart).
| Status | established: shown on many cases, with a control. |
| Devices | Solar8000 monitor |
What is done about it¶
The TimeNet import drops a record that repeats the previous value within 50 ms, and counts the drops.
Evidence¶
| Study | Date | Status | Summary |
|---|---|---|---|
| How the monitors compute their numbers | 2026-10-12 | current | Recomputing each monitor numeric from its own waveform finds a short, simple algorithm for every one (a median, an exponential average or a time mean over 3–30 s) and a delay of 2–9 s behind the newest data it uses. No value is stamped before its data exist. |