Clock links¶
How one clock relates to another: drift and correction. Every rate is relative to against.
| BIS against SNUADC | Measured directly from bursts seen in both the ECG and the EEG, independent of the recorder: −72 ppm (median of 191 benchmark cases), and a constant offset of about −0.8 s left after both stamp models. Per case the slope matches the difference of the two drifts against the PC (Shared bursts between the ECG and the EEG); the release medians of those drifts (+25 and −63 ppm) differ by 88 ppm, being medians over different sets of recordings. |
| BIS against the recorder PC | The BIS EEG stamps drift about +25 ppm against the PC (median from the correction steps of 1,939 recordings; a fit over every recording's stamps gives +21 ppm in 4,728) and are pulled forward on a fixed 8,192 s timer of the BIS's own, which does not start with the recording. |
| CardioQ against SNUADC | Beats matched between the CardioQ's arterial pressure and SNUADC's put the CardioQ on the SNUADC timeline within each re-anchored stretch (matched-beat MAD 0.24 s). |
| CardioQ against the recorder PC | The CardioQ's stamps are re-anchored about every 5 minutes, with jumps of either sign up to ±2 s. |
| Hospital records clock against the recorder PC | The offset between the anaesthesia record and lab systems and the recorder PC cannot be measured in the release: it is confounded with documentation lag. Entries rarely precede the device events they document. |
| Monitor output clock against SNUADC | The 240 Hz line in the monitor's analogue outputs runs −25 to −29 ppm against the SNUADC sample clock (6 files). |
| Oximeter frames against SNUADC | The oximeter's 33.3 ms frame period is fixed per setup against the SNUADC sample clock, +7 to +92 ppm across 42 PC keys, and keeps its phase through the sawtooth slips. |
| PPG content against the oximeter frames (the sawtooth) | The PPG content runs 18–52 ppm against the oximeter's frames, fixed per setup. When a whole frame of lag has built up, one frame is repeated: the delay steps up by 33.3 ms, then creeps down again. |
| Pacemaker against SNUADC | At a fixed pacing rate the pacing interval, measured on the SNUADC sample count, agrees between stretches to 0.05–0.1 ms: the SNUADC sample stream is continuous apart from the resets and gaps. |
| Primus against the recorder PC | The Primus stamps drift a median +104 ppm against the recorder PC (measured by breath matching through the SNUADC stamp model). Its small corrections every 7,230 s, each about 21 s after a pause in the AWP waveform, do not keep up with that drift. |
| Recorder PC against UTC | The recorder PC keeps Windows time, synchronised to the hospital's time server every 8–24 hours. Its offset from UTC is assumed, not observed; no step of 0.5 s or more appears anywhere in the release. |
| SNUADC (frequent-correction mode) against the recorder PC | In frequent-correction mode the SNUADC stamps follow the PC, with small forward steps; the −0.95 ppm they show against the sample count is stamp arithmetic. |
| SNUADC (step mode) against the recorder PC | In step mode the SNUADC stamps drift by tens of ppm and are pulled back by about half a second whenever the drift reaches that size, about every two hours. |
| Solar8000 numerics against SNUADC | The Solar8000's numbers carry no clock of their own in the release; their delay behind the SNUADC waveforms they summarise is fixed within a case, which bounds any drift of the monitor's timing. |