Signal lost at a SNUADC step-mode correction¶
Signal time that is never sampled at each step-mode correction, on the sample count.
| Value | 17.7 ms (range 17.2 to 18.2; n = 3; 3 corrections in 2 paced cases; ECG–EEG bursts give 13.0 ± 1.7 ms) |
| Status | measured: Measured in a study. |
| From | Paced ECG as an independent clock |
| About | SNUADC (step mode) against the recorder PC, ECG lead II (SNUADC) |
| Date | 2026-10-17 |
Notes¶
Adopted in the stamp model as 17.7 ms. A refit with this loss left a burst-lag step of −3.3 ± 1.8 ms, which suggests the average loss may be nearer 15 ms.