Three ART tracks carry the PLETH conversion factors

In case 818, case 5175 and case 5693 the SNUADC/ART track holds arterial pressure, but its header gives the PLETH conversion (0.394983 per code, offset −198.2504) instead of the ART one (0.987457 mmHg per code, offset −495.626). Converted as stored, the pressure reads about 2.5 times too low.

Status established: shown on many cases, with a control.
Devices SNUADC
Variables Arterial pressure (SNUADC), Plethysmogram (SNUADC)

What it means

Read with the stored factors, the median arterial pressure in these cases is 28, 25 and 36 mmHg; read with the ART factors it is 69, 62 and 89 mmHg, matching the monitor's own mean arterial pressure (Solar8000/ART_MBP: 70, 63 and 87 mmHg). The raw codes are an ordinary arterial waveform (medians 572, 565 and 592, against 578 for ART over the release). None of the three cases has a SNUADC/PLETH track.

What is done about it

The TimeNet import converts these three tracks with the ART factors (code × 0.987457 − 495.626 mmHg) by default (VITALDB_SCALE_FIX=on) and keeps the stored factors in the metadata. With other readers, convert them by hand. Across the release, the Solar8000 ART numerics are 2.35–2.38 × the waveform read with the stored factors, as expected for arterial pressure stored with the PLETH factors. Found by the signal survey (every conversion factor of every track).