BIS index¶
The BIS depth-of-anaesthesia index, computed from a window of EEG.
Stub
Imported from the time-source table; the full page is still to be written.
| Name in the dataset | BIS/BIS |
| Type | numeric |
| Device | BIS monitor |
| Stamped by | BIS clock |
Measurement space¶
Time side (as stamped): the true time lies between -65 s and -3 s from the stamp: the delay range plus the interval one value describes.
Time¶
From UTC to the stored time stamp¶
Derived from this signal's timing table; a written chain is still to come.
Timing¶
Against the working reference. Delay: from the end of the interval a value describes to its stamp. Interval: how wide that interval is.
| Form | Delay | Interval | Clock domain | Status | Evidence |
|---|---|---|---|---|---|
| as stamped (default) | 3 s – 4 s | 16 s – 61 s | BIS clock | measured | 3-4 s after its EEG window ends; window 16 s (48%) or 32-61 s (40%); half the weight older than 30 s |
Combine it with other signals in the synchronisation planner.
Delays¶
- From the EEG window end (3 s (p10–p90 3 to 4)): The same in every smoothing group.
Averaging window¶
Trailing average (16 s in 48% of good fits (the 15 s setting); 32–61 s in 40%; half the weight older than 30 s): Set per case on the monitor; not recorded.
Value¶
Its value side is still to be described.
Parameters¶
| Parameter | Value | Status | From |
|---|---|---|---|
| BIS index delay | 3 s (p10–p90 3 to 4) | measured | How the BIS index trails its EEG |
| BIS index averaging window | 16 s in 48% of good fits (the 15 s setting); 32–61 s in 40%; half the weight older than 30 s | measured | How the BIS index trails its EEG |
Observations¶
| Observation | Status | Summary |
|---|---|---|
| Monitor numerics describe the past, never the future | established | Every monitor numeric describes a window of 3–30 s or more that ended 1–9 s before its stamp. None is stamped before its data exist, so the stamps already respect the availability rule. |
| The BIS index is slow | established | The BIS index averages 15–60 s of EEG ending about 3 s before its stamp, with half its weight older than 30 s. It cannot time anything faster than about a minute. |
Studies¶
| Study | Date | Status | Summary |
|---|---|---|---|
| Value resolution of every signal | 2026-10-19 | draft | How finely every waveform and numeric track in the release records its values. The SNUADC waveforms are 10-bit codes, though VitalRecorder's documentation describes the SNUADC as 12-bit; Primus waveforms use about 10 bits of 12; BIS EEG is a 16-bit grid of 0.05 µV. Numerics are 32-bit floats on decimal grids of 1, 0.1, 0.01 or 0.001, one per track; settings sit on coarser grids of 10 or 100. |
| How the BIS index trails its EEG | 2026-10-13 | current | Modelling the BIS index as a lagged mix of EEG features shows a trailing average starting about 3 s after its EEG window ends: about 16 s long in half the cases (the 15 s smoothing setting) and 32–61 s in most others. Half the response weight is older than 30 s. |
| Drug effects as a clock check | 2026-10-13 | current | Infusion pumps, the rapid infuser and the cerebral oximeter share no waveform with the reference. Drug effects that follow infusions at plausible delays (propofol to BIS in about 2 minutes, fluid to CVP in about 30 s) bound their clocks to well within a minute of the recorder. |
| What a BIS-mimic project says about the index's smoothing | 2026-10-13 | current | The openeeg project reproduces the BIS index from the raw EEG and infers each case's smoothing setting: about 80% of its cases come out at 15 s. Its alignment cannot serve as a timing reference. |
| Anaesthesia record times are rounded up | 2026-10-12 | current | Anaesthesia start and end entered on a 5-minute mark sit 90–290 s later, against device markers, than entries typed to the minute. Rounding up to the next 5 minutes fits every one of 11 markers better than rounding to the nearest or down. |
Worked examples¶
| Example | Side | Status | Summary |
|---|---|---|---|
| Post-induction hypotension, from the charted start and from the devices | time | draft | The charted anaesthesia start comes a median 14 minutes before the patient is induced; measured from it, "hypotension within 20 minutes of induction" is found in 27% of 2,913 patients, against 51% from the devices' induction, and the label changes for a quarter of patients. |
Tags: stub