Monitor analogue output clock¶
A line at 240 Hz in the arterial pressure and plethysmogram (weakly in the ECG) runs −25 to −29 ppm against the SNUADC sample clock: most likely the patient monitor's output stage.
Stub
Imported from the time-source table; the full page is still to be written.
| Type | oscillator |
| Nominal | 240 Hz line in the analogue outputs |
Route to UTC¶
The route the atlas trusts most (3 links; its weakest link is estimated apart from the link to UTC that every clock shares):
link measured SNUADC (frequent-correction mode) against the recorder PC: drift −0.954 ppm (n = 6321; census of 6,321 files (17 October): median −0.9535 ppm; 89.5% of recording hours within 0.05 ppm of −0.954; 747 files (12%) between −0.90 and −0.55 ppm)
Where every route ends.
Independent checks. 3 other routes reach UTC; where they meet, the offsets must close around the loop:
- Monitor analogue output clock (Monitor output clock against SNUADC) → SNUADC sample clock (SNUADC (step mode) against the recorder PC) → Recorder PC clock (Recorder PC against UTC) → UTC
- Monitor analogue output clock (Monitor output clock against SNUADC) → SNUADC sample clock (BIS against SNUADC) → BIS clock (BIS against the recorder PC) → Recorder PC clock (Recorder PC against UTC) → UTC
- Monitor analogue output clock (Monitor output clock against SNUADC) → SNUADC sample clock (CardioQ against SNUADC) → CardioQ clock (CardioQ against the recorder PC) → Recorder PC clock (Recorder PC against UTC) → UTC
Plan. Measured against SNUADC on 6 files only. To strengthen: run the line scan on all cases, settle whether the line is 240 Hz or an aliased 260 Hz (which flips the sign), and match it to the PC-stamped Solar8000 numerics for a second route.
Links to other clocks¶
| Link | Clock | Against | Modes | Rate | Corrected every | Status |
|---|---|---|---|---|---|---|
| Monitor output clock against SNUADC | Monitor analogue output clock | SNUADC sample clock | all | −25.4 to −28.7 ppm (6 cases) | — | estimated |
Parameters¶
| Parameter | Value | Status | From |
|---|---|---|---|
| Monitor 240 Hz line against SNUADC | −25.4 to −28.7 ppm (6 cases) | measured | Monitor clocks in the waveforms' spectral lines |
Studies¶
| Study | Date | Status | Summary |
|---|---|---|---|
| Monitor clocks in the waveforms' spectral lines | 2026-10-17 | open | Differences of the raw converter counts show lines from the monitor's analogue outputs: 240 Hz in arterial pressure and plethysmogram, and a 30 Hz family in the plethysmogram alone. Each measures a monitor clock against SNUADC. |
| The SNUADC firmware | 2026-10-13 | current | The public SNUADC firmware explains the board's timing: it sends no time stamps, samples on its 16 MHz crystal, loses time rather than bunching samples after a stall, reads its channels one after another, and truncates its values. A predicted fault at the microsecond counter's wrap was not found in the data. |
Tags: stub